Status
Conditions
About
This study aims to evaluate the Zenalux IM1, an optical spectroscopy device with a pressure sensor. The pressure sensor is intended to overcome the inconsistencies attributable to variations in use of the device. This study will involve using the device on healthy volunteers. Three tissue parameters will be used to analyze the pressure effects in two ways: within- and across-subject variability. Measurements will be taken under the tongue and on the inside of the cheek. The device is noninvasive and is approved for preclinical use.
Full description
PURPOSE OF THE STUDY:
STUDY OBJECTIVES
DESIGN AND PROCEDURES:
Evaluation of the pressure sensor in a volunteer study (n=10)
Informed consent will be obtained from all volunteers. In order to assess the pressure sensor, all research procedures will be repeated twice (once with the aid of the stepper motor and once without the aid of the stepper motor.) For the first five subjects, measurements will be made with the stepper motor first followed by measurements without the stepper motor. The remaining five-subjects will receive the measurements in the reverse order (no stepper motor following by stepper motor). Before the optical measurements, all subjects will be asked to rinse their mouth with a 0.9% saline solution in order to minimize the influence of consumed food. The instrument is high-level disinfected prior to each use in cidex or sterilized by ethylene oxide per standard clinical procedure. Furthermore, it is not anticipated that this device will pose a significant risk to vulnerable populations. The probe will be covered with an optically clear protective sterile sheath and placed in contact with the soft tissue. Reflectance measurements will be taken from each volunteer at up to 5 pressure levels (25, 50, 75, 100, 150 mmHg). At each pressure level 10 spectra will be acquired sequentially. The first measurement will start immediately after the desired probe pressure is reached. This will generate a total of 100 sample and calibration spectra pairs from each subject. This will be repeated on other sites including the tongue, cheek and gums. The sample spectrum will be calibrated using the calibration spectrum collected concurrently. Three tissue parameters will be used to analyze the pressure effects which is described here with [THb] as an example: First, the mean [THb] and its standard deviation will be calculated from the 10 repeat measurements at each pressure for each subject for measurements with and without the stepper motor and plotted against the probe pressure (refer to as THb-P curve). Second, at each pressure level, the [THb] will be plotted against time for each subject and the curve will indicate the temporal response of the tissue under that probe pressure. A series of paired t-tests will be used to identify the contact pressure at which minimal [THb] variance occurs over time (when stepper motor is used). Next, the variance at that contact pressure will be compared to the case without the stepper motor to demonstrate a significant reduction in pressure variance.
The investigators do not anticipate any risks to participants, however, there are a few potential minimal risks such as loss of confidentiality; slight risk of infection; discomfort at the pressure sites; and jaw discomfort as a result of the mouth being opened for several minutes.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Loading...
Data sourced from clinicaltrials.gov