The purpose of this study was to examine the validity of the measurement method based on observation of vibration stimulation-MUAP (motor unit action potential) (1 to 1) matching that helps to measure human WBV-induced muscular reflex latency and to compare these results with the latency determination obtained using the classical methods. The investigators' principle hypothesis is that the simple method is capable of determining the WBV induced reflex response latency as reliably as the complex methods described earlier.
To achieve this goal, the investigators have averaged the electrical responses of the soleus muscle (sEMG and multi motor unit (MMU) potential) against the vertical force data obtained during all the six vibration frequencies tested in this study. The investigators have then combined all the averaged EMG data belonging to six different vibration frequencies to generate the 'cumulative averaged' sEMG and MMU for each subject. The investigators have also obtained the standard error for the cumulative averaged records which indicated the WBV-induced muscular reflex (WBV-IMR) latency.
Full description
Subjects Five healthy volunteers will be participated in the study.
Procedures During WBV, the subjects will be asked to stand upright on a vibration platform (PowerPlate Pro5, London, United Kingdom). Their hips and knees will be in a neutral position. Subjects will be barefooted, and no sponge or foam will be placed between the vibration platform and their feet (Figure 1). The whole plate oscillates with a linear movement upward and downward. Vibration amplitude is 2.2 mm. WBV frequencies will be varied randomly to cover the following six frequencies: 25, 30, 35, 40,45 and 50 Hz. Each frequency lasts for 60 seconds and there will be a 15 seconds rest in between different frequencies.
A load sensor (FC2331-0000-2000L Compression Load Sensor, France) will be fixed on the WBV platform in order to be able to determine induced muscular response latency. Subjects will be positioned on the WBV platform such that their right heel will be placed on the load sensor.
Recordings MMUs, surface EMG (sEMG) and force recordings will be simultaneously done using the Powerlab data acquisition system (ADInstruments, Oxford, United Kingdom). A sampling frequency of 40 kiloHertz (kHz) will be used. Data will be analyzed offline using the MATLAB (R2012a 7.14.0.739).
In order to record sEMG, Ag/AgCl electrodes [KENDALL ® Arbo] with a disc radius of 10 mm will be placed 20 mm apart on the right soleus muscle belly, on shaved skin cleaned with alcohol. The ground electrode will be placed on the lateral malleolus.
To record MMU activity, a pair of Teflon® insulated silver wires (75 µm in core diameter) will be used. About 3 mm of the tips of the silver wires will be stripped of their Teflon ® coating in order to be able to record the activity of several motor units. The pair of wires will be inserted into the right soleus muscle with a 25 G needle. Needle will be immediately withdrawn leaving the fish-hooked wires in the muscle. A lip-clip electrode will be used as the ground electrode.
All cables will be carefully taped to the skin to minimize mechanical artifacts.
The subjects will be instructed to relax their muscles throughout the recordings and were trained using EMG feedback to this end. WBV may impair the sense of balance and muscles may be activated to restore balance. To overcome this issue, the subjects will be familiarized with WBV with a 15 s session of WBV, the subjects will be asked to use the handles of the WBV device to secure their balance and the base of support (distance between the heels) will be adjusted as wide. The distance between the heels usually will be about 25 cm.
Filtering and Rectification regimens All force sensor recordings will be 5 Hz high-pass filtered. All sEMG recordings will be 80 - 500 Hz band-pass filtered (Butterworth 1st order) and full-wave rectified to overcome movement artifacts. All MMU recordings will be 100-5000 Hz band-pass filtered (Butterworth 1st order) and full-wave rectified.
WBV - induced muscular reflex (WBV-IMR) latency
In a previous study, right soleus muscle sEMG measurement and force sensor measurement that is placed under right heel during WBV were taken synchronously and it was found that vibration stimulation and MUAPs would be matched 1 to 1. In this study that is based on the previous observation, it is predicted that it is possible to calculate latency by using averaging method.
All MMU and sEMG data will be analyzed by applying respectively filtering, rectification and differentiation with MATLAB (R2012a 7.14.0.739). Data analysis will be performed in two stages: