ClinicalTrials.Veeva

Menu

Automated Applanation Tonometry

Duke University logo

Duke University

Status

Enrolling

Conditions

Glaucoma
Glaucoma, Suspect

Treatments

Device: Upright applanating prototype
Device: Standard GAT
Device: Fixed force GAT
Device: Supine Applanating Prototype

Study type

Interventional

Funder types

Other

Identifiers

NCT05235321
Pro00106111

Details and patient eligibility

About

Goldmann Applanation Tonometry (GAT) is considered the clinical gold standard for eye pressure measurements and yet it is known to be a subjective measurement with limited repeatability and limited portability. The purpose of this study is to develop an automated and objective method for performing applanation tonometry using standard ophthalmic equipment as well as using 2 portable prototypes.

Full description

This is a prospective study. Patients presenting for their scheduled eye appointments will be recruited by verbal communication. Following informed consent, intraocular pressure (IOP) measurements will be made by the following 3 upright methods and 1 supine measurements:

  1. Standard Goldmann Applanation Tonometry (GAT): this is the standard method for IOP measurement in clinical practice. The eye is given topical fluorescein/anesthetic, the GAT prism contacts the eye while the observer looks through the slit lamp machine ocular using blue light illumination to visual the applanation mires. The GAT dial is adjusted until mire alignment is achieved and the IOP measurement is read off the GAT dial.

  2. Fixed force GAT: From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPod touch camera has been previously published by this PI (1).

  3. Upright applanating prototype: With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter.

    The order of the above 3 measurements will be randomized.

    The 4th measurement will be made with the patient in the supine position. Here, the subject will be reclined in our standard clinical exam chair until the iris plane is parallel to the ground.

  4. Supine applanating prototype: With this method, a 5 gram clear acrylic cylinder (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is aligned with the lens of the CMOS camera and the distal tip of the cylinder is illuminated with blue light using an LED similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. The eye is given topical fluorescein/anesthetic. While the CMOS camera is recording, the 5 gram weight will rest upon the eye and circular applanation mires are recorded. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter. A similar method using an iPhone 6 camera has been previously published by this PI (2). The clear cylinder in this study is the same as that used in this prior study.

Enrollment

200 estimated patients

Sex

All

Ages

18+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

Presenting for a routine eye exam

≥ 18 years of age Able and willing to give consent

Exclusion criteria

History of corneal scarring Active infection of the eye

Trial design

Primary purpose

Diagnostic

Allocation

Randomized

Interventional model

Parallel Assignment

Masking

None (Open label)

200 participants in 4 patient groups

Standard GAT
Active Comparator group
Description:
This is the standard method for IOP measurement in clinical practice
Treatment:
Device: Standard GAT
Fixed-force GAT
Experimental group
Description:
From the patient perspective, this method will feel identical to the standard GAT. The eye is given topical fluorescein/anesthetic. The GAT dial is set at 1.8 or 2.0, a C-MOS camera is connected to one of the oculars of the slit lamp machine and under blue light illumination, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance through the ocular. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter
Treatment:
Device: Fixed force GAT
Upright applanating prototype
Experimental group
Description:
With this prototype, an applanating prism (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is attached to a fixed-force spring that creates a force equivalent to 1.8 or 2.0 on the GAT dial. Blue LED lights on the prototype are used to create the blue illumination similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. A C-MOS camera is aligned with the GAT prism to image the applanation mires. The eye is given topical fluorescein/anesthetic. Like fixed-force GAT, the GAT prism contacts the eye while the CMOS camera makes a video of the mire appearance. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter
Treatment:
Device: Upright applanating prototype
Supine Applanating Prototype
Experimental group
Description:
With this method, a 5 gram clear acrylic cylinder (custom manufactured with medical grade acrylic in an ISO-13485 certified facility) is aligned with the lens of the CMOS camera and the distal tip of the cylinder is illuminated with blue light using an LED similar to the blue light used in clinical practice on the slit lamp or Perkins tonometer. The eye is given topical fluorescein/anesthetic. While the CMOS camera is recording, the 5 gram weight will rest upon the eye and circular applanation mires are recorded. The diameters of the recorded mire images are measured and the IOP is calculated based on the mire diameter
Treatment:
Device: Supine Applanating Prototype

Trial contacts and locations

1

Loading...

Central trial contact

Sarah Jones, MS

Data sourced from clinicaltrials.gov

Clinical trials

Find clinical trialsTrials by location
© Copyright 2025 Veeva Systems