Status
Conditions
Treatments
About
In post-dilution haemodiafiltration only synthetic membranes have been used to date.
The allergy problems described with these membranes require the development of other membranes capable of performing this treatment. We describe in vivo performance and behaviour of an asymmetric cellulose triacetate(ATA™) membrane, to identify its depurative effectiveness and ease of use in clinical practice, as well as evaluate its biocompatibility in a single haemodialysis session (acute biocompatibility) and after one month of treatment (chronic).
Full description
An interventional study was performed in 3 hemodialysis hospital units (La Princesa University Hospital, Príncipe de Asturias University Hospital and Infanta Leonor University Hospital, Community of Madrid, Spain) in which the usual dialyzer that each patient had for OLHDF was replaced with the Solacea™ dialyzer keeping the rest of the parameters unchanged. The study (LIB 09/2015) was reviewed and approved by the CEIC of the Príncipe de Asturias University Hospital.
Design of the study Each patient underwent 12 sessions of haemodialysis with the usual schedule and monitor: 5008 of Fresenius (n=14), AK200US (n=5) and Artis of Gambro (n=3) and DBB007 of Nikkiso (n=1) all suitable for OLHDF, although with different convective transport control systems. The nursing staff connected the automatic OLHDF system. In the case where the system was Ultracontrol® (Gambro monitors) if alarm of PTM> 300 mmHg or system pressure (PSist)> 700 mmHg appeared and were not solved, Ultracontrol® would be withdrawn and the pressure-control system would be used.
Data to be collected:
Demographic and dialysis data
The filtration fraction (FF) was calculated as the percentage of Qi relative to Qb. The convective volume (Vconv) was defined as the total ultra-filtrated volume, which is the sum of VI and UF.7
Analytical determinations
Blood
In CI and CP the following parameters were measured: haemoglobin, proteins and albumin, urea, phosphorus, creatinine, uric acid, β2-microglobulin, myoglobin and retinol transport protein (RTP).
Leukocytes, platelets, C3a and C5a were quantified in CI and CM.
--- Laboratory determinations
General biochemical data: haemoglobin, proteins, albumin, urea, phosphorus, reactive protein C (PCR), creatinine and uric acid, β2-microglobulin, myoglobin, and (RTP) were determined with the usual analyser of each hospital.
Determinations of monocytes, complement, IL-6 and IL-1β were performed in the laboratory of the University of Alcalá:
Activation of plasma complement using a sandwich ELISA: Determination of C3a and C5a activation was performed on platelet rich plasma samples using commercially available ELISA kits C3a Elabscience (Wuhan, P.R. China) and ELISA C5a RayBiotech (Norcross, Georgia).
Determination of interleukin concentration in serum using a sandwich ELISA: IL-6 and IL-1β concentrations were measured in serum samples obtained from whole blood and stored at -80 ° C until use. Both kits used were supplied by Abcam (Cambridge, UK)
Determination of monocyte subpopulations: The different subpopulations of circulating monocytes in peripheral blood were identified by flow cytometry (FACSCalibur™, Becton Dickinson, San Jose, CA, USA). First of all, the monocyte population was selected according to size (FSC/forward scatter) and granularity (SSC/side scatter) and then the different subpopulations were selected by double immunofluorescence according to staining with the anti-CD14 conjugated Tricolor (monoclonal antibodies TuK4) and anti-CD16 conjugated FITC (monoclonal antibodies 3G8). Both antibodies and their respective isotype controls were purchased from Life Technologies Invitrogen (California, USA). The analysis was performed with the Cyflogic program.
For protein bound substances and β2-microglobulin concentrations at the end of the session were corrected for haemoconcentration by a correction factor (FC) based on plasma protein concentration (PT):
FC = PTpre / PTpos, where PTpre and PTpos are the total concentration of proteins pre-dialysis and post-dialysis.
--- Statistical analysis All data was collected in a database (SPSS version 15). Each value was obtained with the average of the values obtained in the different sessions or analytical determinations.
For the statistical analysis, descriptive tools were used, showing the average (standard deviation), median, quartiles or percentages as appropriate. For the comparison of two independent continuous variables, the Student t test for paired samples was used. For the comparison of more than two quantitative variables the ANOVA test was used. The p <0.05 was considered statistically significant.
Enrollment
Sex
Ages
Volunteers
Inclusion and exclusion criteria
Inclusion
Exclusion Criteria:
Primary purpose
Allocation
Interventional model
Masking
Loading...
Data sourced from clinicaltrials.gov
Clinical trials
Research sites
Resources
Legal