20 volunteer test subjects will be entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the Polso Monitoring system
Full description
20 volunteer test subjects will be entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the Polso Monitoring system.
A range of stable respiratory rates will be elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers.
Each subject will be instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor will be used as the Accuracy Reference Device (Reference) in this study.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Subject is considered as being morbidly obese (defined as BMI >39.5)
Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study.
Subjects with known respiratory conditions such as:
Subjects with self-reported heart or cardiovascular conditions such as:
Self-reported health conditions as identified in the Health Assessment Form
Cancer / chemotherapy
Other known health condition should be considered upon disclosure in health assessment form
Primary purpose
Allocation
Interventional model
Masking
20 participants in 1 patient group