According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co., Ltd., the clinical group and evaluation criteria presented in, the effectiveness of blood pressure measurement performance was evaluated.
The auscultation method used as a reference (gold standard) of a non-invasive blood pressure (NIBP) measurement method was used. The first blood pressure measurement was performed once each (reference BP not used for evaluation, R0; test BP not used for evaluation, T0) with auscultation and the test medical device (BM5 equipped with BN1 (BNiBP)). The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each blood pressure measurement.
Three data pairs (mean value of R1, R2 & T1 / R2, mean value of R3 & T2 / R3, mean value of R4 & T3) were obtained from one study subject. This study aims to verify that the mean value of the differences and standard deviations of data pairs meet the criteria of ISO 81060-2:2018.