The clinical protocol of the clinical testing of this device:
Full description
Hospital Information The data was collected by West China Medical College Clinical Investigator Team at Wuhou District Hospital, No. 9 Tongzilin Road, Wuhou District, Chengdu 610041, P.R. China.
Investigator: Catharine Zhang, Nurse A; Vivian Luo, Nurse B; Dr. James Zhu, Sponsor.
Contact Dr. James Zhu Tel: +8613308036568 Used Equipments DUT (Device Under Test): Transtek Blood Pressure Monitor, TMB-986 Reference Device: Yuyue medical BP meter, BP212, accuracy: ±1mmHg and range: 0-300mmHg.
Test Protocol
Test Purpose:
The aim of clinical test is to determine, for the purpose of design qualification, the overall performance of the system meeting the following requirement: For systolic and diastolic pressures, treated separately, the mean difference of the paired measurement of the test system and the comparison system shall be ± 5mmHg or less, with standard deviation of 8mmHg or less.
Target Subject:
The clinical study was conducted according to the protocol described in the ANSI/AAMI SP10 with 95 subjects.
Test procedures (auscultatory method, same-limb sequential measurement was chosen)
Record the age, gender, and circumference of the left arm of the subject.
Let subject seated or supine calmly for two minutes before test.
Measure heart rate in one minute by palpation on radial artery, record it.
Fig. 0 Method of auscultatory validation 4) As shown in Fig. 0, two observers shall make simultaneous, blinded, blood pressure determination on the subject's left arm. Each nurse will write down the reading on a small piece of paper, make a tick if there is an auscultatory gap and then submit to the recorder.
The two readings separately from two observers will be recorded, and if two observers agree that they find an auscultatory gap this time, it will also be recorded. The arm cuff will be take off, after 90 seconds rest, TMB-986 cuff will be taken on and tested, and the readings (systolic, diastolic, mean BP and Heart rate) be recorded.
For each subject, repeat 3) and 5) to get 3 measurements. 7) Repeat step 1) to 6) for every subject. 5. Note: