L

LumiraDx

Status

Not yet enrolling

Conditions

Influenza A
Influenza B
SARS-CoV-2
Asymptomatic COVID-19

Treatments

Diagnostic Test: LumiraDx SARS-CoV-2 & Flu A/B

Study type

Observational

Funder types

Industry

Identifiers

NCT05787444
13379 (Other Identifier)
CS-1489-01

Details and patient eligibility

About

This is a prospective multicenter study conducted to evaluate the performance of the LumiraDx SARS-CoV-2 & Flu A/B tests at point of care sites. Subjects presenting with symptoms suggestive of coronavirus disease 2019 (COVID-19) and Influenza at the time of the study visit will be enrolled and asked to donate swab sample(s) for testing on the device under evaluation.

Full description

This will be a prospective study at point of care sites in the United States for the clinical validation of the LumiraDx SARS-CoV-2 & Flu A/B test for the differential detection of the SARS-CoV-2, and influenza A & B using anterior nasal samples. Up to 1,500 or more symptomatic subjects who are currently experiencing symptoms associated with COVID-19 and Influenza A & B may be enrolled.

The samples are obtained from each subject using standard collection methods and are processed and tested by untrained operators without laboratory training and with no prior experience using the candidate test.

The primary objective of this study is to evaluate the performance (positive percent agreement, negative percent agreement and 95% confidence intervals for upper and lower bounds) of the LumiraDx SARS-CoV-2 & Flu A/B test for the qualitative and differential detection of SARS-CoV-2, Influenza A & B virus antigen using anterior nares samples collected by a healthcare professional, compared to a 510(k) cleared or Emergency Use Authorized (EUA) high-sensitivity reverse transcription-polymerase chain reaction (RT-PCR) test for COVID-19 and a 510(k) cleared Influenza A/B test.

The secondary objective is to assess the usability and user accuracy/performance around limit of detection (LoD) of the investigational test using a questionnaire and accuracy study.

Enrollment

1,500 estimated patients

Sex

All

Ages

2+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  1. Written informed consent obtained prior to study enrollment.
  2. Male or female aged 2 years of age or older
  3. Subject is currently exhibiting fever, or one or more symptoms associated with COVID-19 or influenza (such as, but not limited to, chills, cough, shortness of breath or difficulty breathing, fatigue, new loss of taste or smell, sore throat, congestion or runny nose, nausea or vomiting or diarrhea) and must present within 6 days of symptom onset. Subject must still be exhibiting symptoms on the day of sample collection.

Exclusion criteria

  1. Subject does not understand or is not able and willing to sign the study informed consent or be verbally assented.
  2. Subject has had the seasonal influenza and/or the SARS-CoV-2 vaccine within the past 5 days.
  3. Subject is not able to tolerate sample collection.
  4. Subject is currently undergoing antiviral treatment such as baloxavir marboxil (trade name Xofluza®), oseltamivir (Tamiflu®), zanamivir (Relenza®), and peramivir (Rapivab®).
  5. Subjects undergoing treatment currently and/or within the past thirty (30) days with prescription medication to treat novel Coronavirus SARS-CoV-2 infection, which may include but is not limited to Remdesivir (Veklury), Paxlovid, molnupiravir or receiving convalescent plasma therapy for SARS-CoV-2.
  6. Subjects who have taken Evusheld (tixagevimab/cilgavimab) within the last 90 days.
  7. Subjects who have had a nasal wash or aspirate as part of their standard of care treatment on day of study visit prior to study sample collection.
  8. Subjects who have had recent craniofacial injury or surgery, including to correct deviation of the nasal septum, within the previous six (6) months.

Trial design

1,500 participants in 1 patient group

symptomatic
Description:
Subject is currently exhibiting fever, or one or more symptoms associated with COVID-19 or influenza (such as, but not limited to, chills, cough, shortness of breath or difficulty breathing, fatigue, new loss of taste or smell, sore throat, congestion or runny nose, nausea or vomiting or diarrhea) and must present within 6 days of symptom onset. Subject must still be exhibiting symptoms on the day of sample collection.
Treatment:
Diagnostic Test: LumiraDx SARS-CoV-2 & Flu A/B

Trial contacts and locations

1

Loading...

Central trial contact

Philip Estes

Data sourced from clinicaltrials.gov