CME With Different Fluidic Parameters

I

Iladevi Cataract and IOL Research Center

Status and phase

Completed
Phase 4

Conditions

Cystoid Macular Edema Following Cataract Surgery

Treatments

Procedure: microcoaxial phacoemulsification

Study type

Interventional

Funder types

Other

Identifiers

Details and patient eligibility

About

Understanding and modulating fluid parameters is an important, but often overlooked aspect of phacoemulsification. In a previous study we compared the impact of using high fluid parameters versus low fluidic parameters on real-time IOP measured during phacoemulsification. The investigators found that using high parameters resulted in a higher absolute rise in IOP as well as higher fluctuations in the IOP when compared to low parameters. Clinically these higher fluctuations in IOP would translate in a higher chamber instability. Based on the results of this study, the investigators decided to take it further and study the impact of using high parameters (and thus, higher chamber instability) on macular edema and thickness following surgery, in an otherwise uncomplicated surgery.

Higher fluid parameters during phacoemulsification predisposes the eye to increased macular thickness

Full description

Several studies have shown the adverse impact of an increase in the IOP and IOP fluctuations that occur during anterior segment intervention on the posterior segment structures. In human volunteers with each incremental increase in IOP the systolic and diastolic flow velocities in the short posterior ciliary arteries decreased linearly. This implies that the normal healthy eye is not able to autoregulate to maintain posterior ciliary artery blood flow velocities in response to acute large elevations in IOP. Vascular insufficiency due to abnormal autoregulation has been proposed as a major factor in the development of glaucoma. 1

It has been postulated that IOP elevation during the LASIK procedure causes mechanical stress which may induce tangential stress on the posterior segment.2, 3 Some studies have reported that the increase in IOP damages the retinal ganglion cells causing visual field defects. Also sudden increases in IOP, although well tolerated may induce changes in the peripheral retina.4,5,6

Several reports propose the occurrence of macular hole, lacquer cracks and choroidal neovascular membranes following the LASIK procedure. 3 It has been observed that the rapidly fluctuating pressure variations may be detrimental, particularly in susceptible persons with compromised ocular blood flow. Rapid IOP changes across a 30-mm Hg range would be predicted to influence posterior segment blood vessels.

In a previous study we compared the impact of using high fluid parameters versus low fluidic parameters on real-time IOP measured during phacoemulsification. We found that using high parameters resulted in a higher absolute rise in IOP as well as higher fluctuations in the IOP when compared to low parameters. Clinically these higher fluctuations in IOP would translate in a higher chamber instability.

We hypothesize that although transient, the increased IOP that occurs during phacoemulsification when using high parameters could cause mechanical stress on the eye. These higher fluid parameters during phacoemulsification can predispose the eye to increased macular thickness.

To the best of our knowledge there are no published data on impact of IOP changes and fluctuation that are induced during cataract surgery on the macula. To investigate this further, we decided to study the impact of using high parameters (and thus, higher chamber instability) on macular thickness following surgery, in an otherwise uncomplicated surgery.

Enrollment

150 patients

Sex

All

Ages

40 to 70 years old

Volunteers

Accepts Healthy Volunteers

Inclusion criteria

  1. Uncomplicated, Age-related cataract.
  2. Nuclear sclerosis: upto grade 3
  3. Age: 40-70 years
  4. Axial length: 21.5 mm to 24.5 mm

Exclusion criteria

  1. Diabetes mellitus
  2. Co-existing ocular disease- uveitis, glaucoma, PEX
  3. Pre-existing macular pathology (eg.ARMD)
  4. Previously operated eyes
  5. Under treatment with Topical or systemic steroids / NSAID's
  6. Intraoperative complications- PCR, Descemet's detachment, uveal trauma
  7. Post operative complications - severe inflammation (>grade 3), rise in IOP

Trial design

Primary purpose

Prevention

Allocation

Randomized

Interventional model

Parallel Assignment

Masking

Quadruple Blind

150 participants in 3 patient groups

Longitudinal U/S - low fluidic
Active Comparator group
Description:
ASPIRATION FLOW RATE - 25 CC/MIN, BOTTLE HEIGHT - 90 CMS, LONGITUDINAL ULTRASOUND
Treatment:
Procedure: microcoaxial phacoemulsification
Torsional U/S - low fluidic
Active Comparator group
Description:
ASPIRATION FLOW RATE - 25 CC/MIN, BOTTLE HEIGHT - 90 CMS, TORSIONAL ULTRASOUND
Treatment:
Procedure: microcoaxial phacoemulsification
Longitudinal U/S - high fluidic
Active Comparator group
Description:
ASPIRATION FLOW RATE - 40 CC/MIN, BOTTLE HEIGHT - 110 CMS, LONGITUDINAL ULTRASOUND
Treatment:
Procedure: microcoaxial phacoemulsification