Comparison of Intraocular Pressure Measurements Between Reichert Tono-Vera Tonometer and Goldmann Tonometry

R

Reichert, Inc.

Status

Completed

Conditions

Glaucoma
Intraocular Pressure

Treatments

Device: IOP with Tono-Vera Tonometer
Device: IOP with comparator ORA G3 and ic100 tonometers
Device: IOP with Goldmann Applanation Tonometer

Study type

Interventional

Funder types

Industry

Identifiers

NCT04360369
16305-TPR-28

Details and patient eligibility

About

Determine if the Tono-Vera Tonometer accurately measures intraocular pressure (IOP). The hypothesis of this test is to confirm the Tono-Vera Tonometer is equivalent to the Goldmann Applanation Tonometer (within +/- 5.0 mmHg).

Full description

The Tono-Vera Tonometer (Reichert, Inc.) is a portable rebound tonometer intended to measure IOP. Subjects will undergo a standard ophthalmologic examination. Subjects will then be measured with four tonometers: Goldmann Applanation Tonometer (C.S.O. SRL), Ocular Response Analyzer G3 (Reichert, Inc.), iCare ic100 tonometer (iCare Finland Oy), and Tono-Vera Tonometer.

Enrollment

61 patients

Sex

All

Ages

18 to 90 years old

Volunteers

Accepts Healthy Volunteers

Inclusion criteria

  • Subjects must be male or female, between the ages of 18 and 90 years old;
  • Be able and willing to provide signed informed consent
  • Be able to follow study instructions

Exclusion criteria

  • Subjects with only one functional eye;
  • Subjects with one eye having poor or eccentric fixation;
  • Subjects with central corneal thickness greater than 600 µm or less than 500 µm (about 2 standard deviations the human mean);
  • Subjects with corneal scarring or who have had corneal surgery, including corneal laser surgery;
  • Subjects with concomitant ocular diseases such as: microphthalmos, buphthalmos, nystagmus, keratoconus, severe dry eye syndrome, blepharospasm, any other corneal or conjunctival pathology or infection;
  • Contact lens wearers;
  • Known allergy to proparacaine or fluorescein as these are used to anesthetize the eye and allow IOP measurement, respectively, when used with the GAT.

Trial design

Primary purpose

Diagnostic

Allocation

Non-Randomized

Interventional model

Sequential Assignment

Masking

None (Open label)

61 participants in 3 patient groups

Goldmann Applanation Tonometer
Active Comparator group
Description:
Measurement of IOP with Goldmann Applanation Tonometer. All subjects will participate in this arm.
Treatment:
Device: IOP with Goldmann Applanation Tonometer
ORA G3 and ic100
Active Comparator group
Description:
Measurement of IOP with Ocular Response Analyzer G3 and ic100 tonometers. All subjects will participate in this arm.
Treatment:
Device: IOP with comparator ORA G3 and ic100 tonometers
Tono-Vera Tonometer
Experimental group
Description:
Measurement of IOP with Tono-Vera Tonometer. All subjects will participate in this arm.
Treatment:
Device: IOP with Tono-Vera Tonometer