Confirm Rx™ Versus Reveal LINQ™ - Which is More Reliable in Data Transmission? A Randomized Clinical Study

S

Sparrow Clinical Research Institute

Status

Completed

Conditions

Atrial Flutter
Atrial Fibrillation
Syncope
Tachycardia
Stroke
Cardiac Arrythmias

Treatments

Device: Abbott, Inc Confirm Rx™ versus Medtronic, Inc Reveal LINQTM

Study type

Interventional

Funder types

Other
Industry

Identifiers

Details and patient eligibility

About

This study will compare the reliability and timeliness in data transmission of the Abbott Confirm Rx™ loop recorder with the Medtronic Reveal LINQ™ loop recorder.

Full description

Implantable cardiac monitor (ICM) is an invaluable tool for diagnosing cardiac arrhythmia (1). Cryptogenic stroke, unexplained syncope and arrhythmia diagnosis are the most common indication for long term cardiac monitoring device. (2, 3). Reveal LINQTM (Medtronic, Inc.) is widely used for all of these purposes in the country. We have previously reported the use of Reveal XT in the use of atrial fibrillation (AF) surveillance in patient's post-AF ablation (2). We indicated that ICM is useful in long term monitoring of atrial arrhythmia in patients at risk of atrial fibrillation. Crystal AF (4) study also showed the importance of using ICM in detecting occult AF in patients with cryptogenic stroke in a timely manner so that appropriate treatment can be prescribed. The key to the success of ICM rest on how timely and reliable the ICM data can be transmitted to physician's office immediately after an arrhythmic events so the data can be analyzed and treatment rendered. Reveal LINQTM data transmission consists of two steps: 1. Data needs to be download into the MyCareLink™ monitor (scheduled daily at 2 am in the morning and monitor needs to be in the within 6 feet proximity of the patient) 2. MyCareLink™ monitor to be connected to the 3 G cellular network for the data transmission. (Fig 1). In cases of patient activated events, data transmission utilizes similar pathway.

We have reported significant delay in data transmission in the Reveal LINQTM occurs frequently and failure to connect among the Reveal LINQTM, MyCareLink™ monitor and the cellular network is the main reason for data transmission disruption and critical data could be concealed for an extended periods. To illustrate the importance of timely transmission of patient data, a Reveal LINQTM was implanted in a 54-year-old gentle man with past medical history of recurrent syncope. On 8/12/2017 the patient developed new onset dizziness and 6 seconds sinus pause was recorded by the device. However, a connection problem of MyCarelinkTM system made data unavailable until next scheduled office visit on 11/15/2017. The physician was notified on the same day and a pacemaker was implanted on the following day, almost 3 months after his significant arrhythmic events.

The newer generation of ICM such as Confirm Rx™ (Abbott, Inc) is connected directly to a patient's existing smartphone using the Bluetooth technology. By allowing patients to record and transmit symptoms with the mobile app, Confirm Rx™ ICM brings continuous remote monitoring to patients without the need for a home-based monitor (Fig 2). It is unclear which technologies offers faster and more reliable transmission of critical patient information to physician office so timely treatment can be provided. We therefore propose a pilot randomized study comparing Confirm Rx™ and Reveal LINQTM in the reliability and timeliness in data transmission.

Enrollment

209 patients

Sex

All

Ages

18+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • Patient is 18 years of age or older
  • Patient is willing and able to provide written informed consent
  • Patient is willing and able to comply with the protocol, including follow-up visits and MyCareLink™ and Merlin™ transmissions (standard of care)
  • Candidates for implantable cardiac monitor

Exclusion criteria

  • Patient is enrolled in another study that could confound the results of this study, without documented pre-approval from a study manager
  • Patient has existing IPG, ICD, CRT-D or CRT-P device
  • Adequate sensing in one week post implant. If sensing not adequate, patient will be in the registry.

Trial design

Primary purpose

Diagnostic

Allocation

Randomized

Interventional model

Parallel Assignment

Masking

None (Open label)

209 participants in 2 patient groups

Abbott, Inc Confirm Rx™
Active Comparator group
Description:
Every other consenting subject will receive the Abbott Inc. Confirm Rx™ device.
Treatment:
Device: Abbott, Inc Confirm Rx™ versus Medtronic, Inc Reveal LINQTM
Medtronic, Inc Reveal LINQTM
Active Comparator group
Description:
Every other consenting subject will receive the Medtronic, Inc. Reveal LINQTM.
Treatment:
Device: Abbott, Inc Confirm Rx™ versus Medtronic, Inc Reveal LINQTM