Status
Conditions
Treatments
About
The purpose of this study is to evaluate the accuracy and reliability of the continuous glucose monitoring system the FreeStyle Navigator® (Abbott, Diabetes Care) in critically ill patients by comparing subcutaneous measuring with the current standard of arterial measuring by a point of care glucometer.
Full description
Most patients admitted to the ICU are treated with intensive insulin therapy, according to a locally developed glucose treatment protocol. The use of a continuous glucose monitoring in critically ill patients could have the following advantages:
In this study, patients will be randomized in: 1. The Freestyle group or 2. The AccuChek Group.
Patients in both groups will receive a Freestyle Navigator device, which will stay on the patient up to 5 days, or until discharge from the ICU.
The medical treatment of both study groups will be equal, except for the frequency of the glucose level measurements and the resulting adjustments according to the treatment protocol. In the Freestyle group extra glucose level measurements will be made in response to alarms from the device.
In both groups, every 4 hours blood glucose values will be obtained using a blood gas analyzer (BGA). When a control measurement by BGA results in a glucose level of <2.2 or >25 mmol/l, the value will be transmitted to the Patient Data Management System (PDMS) and gives an alarm. Otherwise, the value will be blinded in the PDMS.
Freestyle Navigator data will be downloaded from the device. Blood glucose values obtained in the course of usual care will be compared with time-matched Freestyle Navigator values to assess device accuracy.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Primary purpose
Allocation
Interventional model
Masking
178 participants in 2 patient groups