Status
Conditions
Treatments
About
We hypothesize that measurements of interstitial fluid (ISF) glucose by the FreeStyle Navigator (Abbot Diabetes Care), a continuous glucose monitoring system, will correlate with blood glucose (BG) values in surgical and ICU patients with a clinically useful degree of accuracy.
Full description
Critically ill patients treated with intensive insulin therapy will be monitored for up to 72 hours with the Freestyle Navigator continuous glucose monitor. The device display will be disabled so that continuous glucose monitoring results will not be available to care givers, although low (<60 mg/dL) and high (>250 mg/dL) threshold audio alarms will be enabled. Blood glucose monitoring will be performed according to the usual practice for patients treated with intensive insulin therapy in each intensive care unit, except that additional blood glucose measurements will be performed in response to Freestyle Navigator threshold alarm. All blood glucose management decisions will be made according to intensive insulin therapy protocol in each unit. Freestyle Navigator data will be downloaded from the device after each subject has completed the monitoring period. Blood glucose values obtained in the course of usual care, or in response to Freestyle Navigator threshold alarms, will be compared with time-matched Freestyle Navigator values to assess device accuracy and alarm sensitivity and specificity.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
In order to obtain adequate representation of patients with physiological alterations that might decrease correspondence between blood and ISF glucose, we will recruit 50 subjects in each of five groups:
Primary purpose
Allocation
Interventional model
Masking
76 participants in 1 patient group