Status
Conditions
Treatments
About
It has been shown that videolaryngoscopy may be superior to direct laryngoscopy for endotracheal intubation in intensive care. Recently, an endotracheal tube with an integrated camera at its tip has been introduced (VivaSight-SL) allowing for direct visual confirmation of the tube's passage through the vocal cords during intubation.
Patients who are requiring urgent or endotracheal intubation in intensive care are randomized to receive either a conventional intubation with direct laryngoscopy or to receive intubation with the VivaSight-SL-Tube. Primary outcome measures are first attempt success rate and number of attempts to successful intubation.
Full description
Background
Airway management in critically ill patients is usually performed by endotracheal intubation with direct laryngoscopy. [1, 2] However, it has been shown that videolaryngoscopy may be superior to the conventional approach in intensive care, i. e. the number of attempts to successful intubation and the rate of accidental esophageal intubation are lower and the visualization of the vocal chords is improved. [3, 4]
Recently, an endotracheal tube with an integrated camera at its tip has been introduced that permits a continuous visualization of the tube's insertion into the trachea on a monitor connected to the camera (VivaSightTM-SL, ETView Ltd., Misgav, Israel) [5]. This tube has been CE and FDA certified (http://www.etview.com/products/vivasight-sl).
It is possible to observe the passage of the tube through the vocal chords during intubation and to verify the correct placement by visualization of the tracheal cartilage. This is an advantage over videolaryngoscopy, in which the camera is mounted on the laryngoscope blade. For the VivaSightTM-SL tube, a decrease for time to intubation and an increase of first attempt success has been shown over conventional intubation in a simulator trial. [6]
In this study, the VivaSightTM-SL tube is tested against the conventional approach during elective and urgent intubations in critically ill patients.
Methods
Study design: randomized, prospective trial
Sample size: n = 54
With a sample size of 54 (randomized 1:1 as 2x 27) a difference of 35% for first attempt success over conventional intubation may be detected with an α-error of 0,05 and a β-error of 1-0,8.
Duration of study: until sample size reached or 18 months after begin of study
Procedures:
study inclusion:
All patients being treated in the Dept. of Intensive Care Medicine receiving percutaneous tracheotomy due to long term ventilation are screened according to inclusion and exclusion criteria.
Details of study related procedures:
Intubation:
The Intubation with the VivaSightTM-SL endotracheal tube does not differ from an intubation with a conventional tube that is done with respect to the standard operating procedure (SOP) of the Dept. of Intensive Care Medicine. Additionally, the intubation is guided by the camera mounted on the tip of the tube. During the intubation, vital parameters are monitored with respect to the underlying disease and patients' therapy is continuously adjusted. According to the SOP, two physicians are present of which at least one is a fellow or an attending physician with experience in intensive care medicine. In this trial, intubation is done exclusively by a fellow or attending physician.
Consent: all patients or their legal surrogate give written informed consent.
Data protection: Data are anonymized.