ClinicalTrials.Veeva

Menu

Coronary Atherectomy System in Patients With Coronary Calcification (CORECT)

M

MicroPort

Status

Invitation-only

Conditions

Calcification of Coronary Artery

Treatments

Device: Rotablator Rotational Atherectomy System
Device: Coronary atherectomy system of Shanghai MicroPort Rhythm

Study type

Interventional

Funder types

Industry

Identifiers

NCT05447585
AMD-D-001

Details and patient eligibility

About

This is a prospective and multicenter clinical investigation aiming to evaluate the safety and effectiveness of coronary atherectomy system for the treatment of patients with coronary calcification.

Full description

This study is divided into two phases. The first phase is the pilot study stage, which is a prospective single group observational test (FIM stage). The second phase is the pivotal study phase, which is a prospective, multicenter, randomized controlled and non inferiority test (RCT phase). It is carried out in about 25 hospitals in China.

  1. Pilot study phase (FIM phase): It is a prospective, multicenter and single-arm observational test. It is expected to recruit 15 subjects in 3 to 5 research centers in China. The primary endpoint is procedural success rate.
  2. Pivotal study phase (RCT phase): It is a prospective, multicenter, randomized controlled and non inferiority trial. 224 subjects are expected to be recruited in 25 research centers in China and randomly divided into two groups according to the ratio of test group: control group = 1:1. The primary endpoint of the RCT phase is the rate of no major adverse cardiovascular events (MACE) at 30 days after baseline.
  3. All subjects with coronary calcified lesions participating in this clinical study must have a calcified lesion length of no more than 40 mm and located in an coronary artery with a diameter of ≥ 2.50 mm but ≤ 4.00 mm.
  4. All subjects receive clinical follow-up during hospitalization and 30 days, 6 months and 12 months after surgery.
  5. Successful enrollment: The atherectomy guidewires pass through the target lesion successfully (through the true lumen), and there are no vascular perforation, dissection, thrombosis and other complications.

Enrollment

239 estimated patients

Sex

All

Ages

18 to 80 years old

Volunteers

No Healthy Volunteers

Inclusion and exclusion criteria

Clinical Inclusion Criteria:

CI1. Age between 18 and 80.

CI2. Subjects who are going to accept percutaneous coronary intervention (PCI) and drug-eluting stent implantation.

CI3. Asymptomatic evidence of ischemia, stable or unstable angina pectoris, or old myocardial infarction.

CI4. Acceptable candidates for coronary artery bypass grafting (CABG).

CI5. The left ventricular ejection fraction (LVEF) is equal to or greater than 30% measured by echocardiography or left ventricle contrast examination within 30 days before inclusion.

CI6. Subjects who are able to understand the purpose of the trial, participate voluntarily or by proxy and indicate by signing the informed consent form that they recognize the risks and benefits described in the informed consent document and are willing to undergo clinical follow-up.

Angiography Inclusion Criteria:

AI1. There is only one target lesion that needs to be treated.

AI2. The target vessel must be a native coronary artery with reference diameter >= 2.5mm and <= 4.0 mm.

AI3. The target vessel must be with a stenosis of >= 70% and < 100%Target lesion angiographic stenosis, or of >=50% and <70% with local ischemic symptoms.

AI4. Target lesion length ≤ 40 mm .

AI5. The target lesion must have evidence of severe calcium deposit, or moderate calcification which cannot be fully expanded by PTCA balloon catheter at the lesion site based on the protocol criterion, .

Clinical Exclusion Criteria:

CE1. Acute myocardial infarction within 30 days prior to baseline procedure.

CE2. NYHA or Killip class III or IV heart failure.

CE3. Severe cardiac insufficiency with reduced ejection fraction (left ventricular ejection fraction < 30%).

CE4. Intracardiac thrombus within 30 days before inclusion.

CE5. Subjects who have received organ transplantation or is waiting for organ transplantation.

CE6. Subjects who are receiving chemotherapy or scheduled to receive chemotherapy 30 days before or after baseline surgery.

CE7. Subjects who are receiving chronic (≥ 72 hours) anticoagulant therapy (e.g. heparin, warfarin, coumarin) for indications other than acute coronary syndrome.

CE8. Participants' platelet < 50 × 10^9 / L or > 700 × 10^9 / L, and / or hemoglobin < 70g / L.

CE9. Subjects with confirmed or suspected liver diseases, including active hepatitis, or abnormal laboratory results are not suitable for the study.

CE10. Subjects diagnosed with chronic renal failure or has a serum creatinine level >2.5 mg/dl (or 221 µmol / L).

CE11. Subjects who have clear bleeding tendency, contraindications of antiplatelet preparation and anticoagulant treatment, and are unable to receive antithrombotic treatment.

CE12. History of cerebrovascular accident (CVA), transient ischemic attack (TIA), active peptic ulcer disease, gastrointestinal (GI) bleeding, transient ischemic attack (TIA), or permanent neurological defects that may lead to non-compliance with the test protocol in the past 6 months.

CE13. Target vessels (including collateral vessels) received any stent implantation within 12 months before baseline surgery.

CE14. Non target vessels were treated with any type of PCI within 24 hours before baseline surgery.

CE15. PCI or CABG is planned within 2 months after baseline procedure.

CE16. History of coronary endovascular brachytherapy at any time.

CE17. Known hypersensitivity to any drugs required by the stent or protocol (such as rapamycin or structure related compounds, fluoropolymers, aspirin or tigrelol, contrast medium, acrylic acid, stainless steel, chromium, nickel, iron, tungsten).

CE18. Subjects who have other serious medical diseases (such as cancer, congestive heart failure), which may reduce life expectancy to less than 12 months or it is expected to be difficult to complete 12-months follow-up.

CE19. Subjects who currently abuse drugs, such as heroin, alcohol, etc.

CE20. Subjects who plan to undergo surgery that may lead to non-compliance with the protocol or confusion in data interpretation.

CE21. Currently enrolled in any other clinical trial (does not reach its primary endpoint), or intended to participate in any other clinical trial within 12 months after baseline procedure.

CE22. Females subjects who are pregnant,breast feeding or planning to become pregnant within the study period.

CE23. Untrolled diabetes.

CE24. Inability to understand the study or a history of non-compliance with medical advice.

Angiographic Exclusion Criteria:

AE1. Target lesions that meet the following criteria:

  • Ostial lesions (within 5mm of ostium);
  • Target lesions involve bifurcation lesions, and branch vessels need to be implanted with stents;
  • Angular lesions (> 45 °)
  • Located in a native vessel distal to anastomosis with a saphenous vein graft or LIMA/RIMA bypass;
  • Thrombosis in the target vessel;
  • Target lesion dissection;
  • In-stent restenosis;
  • Chronic total occlusion lesions that cannot be passed through by atherectomy guidewire.

AE2. Unprotected left main coronary artery disease (> 50% diameter stenosis).

AE3. Protected left main coronary artery disease (> 50% diameter stenosis at LMCA and left coronary artery receive bypass grafting), and the target lesion is located in LAD or LCX.

AE4. Other clinical significant lesions in target vessels may require intervention within 2 months after baseline surgery.

AE5. The target vessel (including collateral) near the target lesion or within 10 mm (visually) from the distal end of the target lesion has received stent implantation at any time before baseline surgery.

AE6. Especially tortuous target vessels. AE7. Non-target lesion failed to be treated before the treatment of target lesion.

Trial design

Primary purpose

Treatment

Allocation

Randomized

Interventional model

Parallel Assignment

Masking

Single Blind

239 participants in 2 patient groups

Coronary Atherectomy System
Experimental group
Description:
Subjects in experimental arm will be treated with the Coronary Atherectomy System manufactured by Shanghai Microport Rhythm Co. Ltd.
Treatment:
Device: Coronary atherectomy system of Shanghai MicroPort Rhythm
Rotablator Rotational Atherectomy System
Active Comparator group
Description:
Subjects in control arm will be treated with Rotablator Rotational Atherectomy System manufactured by Boston Scientific Corporation
Treatment:
Device: Rotablator Rotational Atherectomy System

Trial contacts and locations

1

Loading...

Data sourced from clinicaltrials.gov

Clinical trials

Find clinical trialsTrials by location
© Copyright 2026 Veeva Systems