This study is designed to measure the trending accuracy of a noninvasive measurement of cerebral oxygen saturation. One Large sensor (adult sensor) and one INVSENSOR00009 will be placed on the subject's forehead. The values obtained by the INVSENSOR00009 will be compared to the values collected with the adult sensor.Data will be collected from healthy adult subjects while undergoing a desaturation procedure wherein the concentration of oxygen inhaled is slowly reduced until the subject's arterial oxygen saturation is approximately 70%. The subject will then be returned to inhaling room air.