This study is designed to measure the trending accuracy of a noninvasive regional oximetry measurement of somatic oxygen saturation. Two large sensors (adult sensors) will be placed on the subject's forehead. Two investigational sensors (INVSENSOR00028) will be placed on the forearm of the volunteer. The values obtained by the INVSENSOR00028 will be compared to the reference value calculated from the arterial and venous blood measurements. Data will be collected from healthy adult subjects while undergoing a desaturation procedure wherein the concentration of oxygen inhaled is slowly reduced until the subject's arterial oxygen saturation is approximately 70%, while blood samples are drawn at prespecified reference points. The subject will then be returned to inhaling room air.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Primary purpose
Allocation
Interventional model
Masking
46 participants in 1 patient group