ClinicalTrials.Veeva

Menu

Effectiveness of Finger and Forehead Plethysmographic Variability Index Monitoring in Pediatric Patients

M

Marmara University

Status

Completed

Conditions

PVI

Treatments

Device: pvi

Study type

Observational

Funder types

Other

Identifiers

NCT04835610
09.2020.132

Details and patient eligibility

About

In pediatric surgical patients, the sensitivity of forehead and finger PVI monitoring was similar for the prediction of fluid responsiveness, whereas the specificity of forehead monitoring was higher. The use of forehead PVI probes in the fluid management of these patients seemed to be appropriate because cephalic region was not affected by low perfusion states.

Full description

he Plethysmographic Variability Index (PVI) can be measured with both finger and forehead probes. Vasoconstriction may reduce the reliability of PVI measurements in pediatric surgical patients prone to hypothermia. The vascular structure of the forehead area is resistant to changes in vasomotor tonus. The aim of our study is to compare the efficacy of PVI values measured at the finger and forehead areas to predict fluid responsiveness in these patients.

Fifty pediatric patients with obtained parental consent and ethics committee approval underwent elective minor surgery were included into the study. Basic monitoring, finger and forehead perfusion index (PI) and PVI monitoring were applied. Hemodynamic parameters, PI, PVI, cardiac output (CO) values were recorded. The 15% increase in CO with passive leg raise (PLR) maneuver was estimated as fluid responsiveness. Two groups were created: Group R (fluid responsive); Group NR (fluid nonresponsive). Mann-Whitney U, t, chi-square and ROC tests were used in statistical evaluation.

Enrollment

2 patients

Sex

All

Ages

5 to 12 years old

Volunteers

No Healthy Volunteers

Inclusion criteria

written consent from their parents patients undergoing elective minor surgery with bleeding less than 500ml Children between the ages of 5-12.

Exclusion criteria

those who stay in the specified age range those with cardiopulmonary disease spontaneous breathing positive allen test peripheral circulatory failure vasopressor therapy anemia hypothermic

Trial design

2 participants in 2 patient groups

PVI values
Description:
PVI values measured at the finger and forehead areas in pediatric patients
Treatment:
Device: pvi
no control group
Description:
no control group
Treatment:
Device: pvi