Status
Conditions
Treatments
About
This investigation will gather information about a procedure called sphenopalatine ganglion (SPG) stimulation, and its appropriateness, safety, and efficacy as a treatment for those who suffer migraine headaches which may result in chronic severe disability. The SPG is a small collection of nerve cells in the head, and is located near the base of the nose on either side. Participation involves the surgical implantation of an electrode (small electrical conductor) over the sphenopalatine ganglion. The electrode is connected to a stimulator which will enable treatment for migraine headaches. Tiny electrical current is delivered to the stimulator device by an internal pulse generator implanted in the area at the top of the chest, to stop the migraine headaches. The implant system will be controlled with a wireless remote provided after the implant procedure.
Participation will record headache diaries throughout the study, which will last approximately 8½ months, and a yearly visit annually for five years.
Full description
This application proposes a clinical study of electrical stimulation of the sphenopalatine ganglia (SPG) as a treatment for up to three individuals with episodic migraine headache. The present study is aimed at obtaining pilot data to guide a future controlled trial of this treatment modality. The study population will include individuals suffering from episodic migraine headaches with chronic severe disability, as demonstrated by the Migraine Disability Assessment Questionnaire (Lipton, 2000) and Headache Impact Test short form (HIT-6™) (Kosinski, 2003).
The treatment involves implantation of an electrode into the SPG. The electrode is connected subcutaneously to an infraclavicular stimulator (PrimeAdvanced™ 37702 Multi-program Neurostimulator System, Medtronic Inc., Minneapolis MN). Proper electrode placement will be verified using anatomic and physiologic techniques. Participants will receive the Medtronic Model 3389 or 3387 lead, or Medtronic subcompact lead Model 3776-45, 3776-60, or 3776-75. Stimulation will be delivered in a range of frequencies from 20 to 130 Hz, and pulse width from 60 to 450 μsec, and a titrated voltage. The voltages used for chronic stimulation may range up to the pulse generator maximum of 10.5 volts but are anticipated to generally be below 3 volts, keeping below the 30 µcoulomb/cm² charge density safety limit, and below the threshold for adverse stimulation-related effects. The minimum number of contacts on the quadripolar leads will be activated as necessary to produce a response.
The Prime Advanced neurostimulator to be used in this study allows the clinician to set all stimulation parameters including the maximum allowable amplitude. It is our responsibility to assure that appropriate stimulation parameters are used to result in appropriate electrical exposure charge density) below a 30 µC/cm2/phase limit. As is done with commercially available neurostimulators, during this study, we will utilize the charts presented in Figures 1 and 2 below to assist in selection of programming parameters, and this programming will be maintained by software controls within the neurostimulator.
For this exploratory study of 3 patients, the maximum electrical stimulating parameters, resulting electrical exposure (e.g., charge density), will be determined using the methods described above. It is our responsibility to assure that safe stimulation parameters are used at all times and that the maximum settings do not exceed the safety limits. Our plan is to begin with lead 3776 if possible, because it does not require an extra extension, but we would use 3387 or 3389 if clinically indicated.
The proposed study is a physician-sponsored research investigation of three patients, and the attention to stimulation parameters will be much greater than can be expected for a commercially released product. It is our responsibility within this investigational study to insure the selected stimulation parameters do not exceed the safety limit of 30μC/cm2/phase. Patients participating in the study will only be able to lower stimulation amplitudes, thereby keeping stimulation parameters BELOW any clinician set maximum. The only programs available to or accessible by the patients will maintain the previously described parameter set limitations. Thus, the programming of the Implantable Programmable Generator (IPG) will keep the three patients from stimulating outside or above the set and safe parameters.
The primary outcome measures for assessing the efficacy of migraine treatment will be a subject reported daily diary noting frequency and intensity of headaches. During this investigation we will obtain preliminary controlled data on the safety and efficacy of SPG stimulation for migraine treatment.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Subject currently has Medication Overuse Headache (MOH) in the judgment of the investigator or by ICHD-II criteria.
Subject has a history of headaches days ≥ 15 per month.
Subject has a history of trigeminal autonomic cephalalgias.
Subject has any medical condition or disorder that:
Subject has had a major infection or surgery in the past month.
Subject has undergone facial surgery in the area of sphenopalatine ganglia for cosmetic, corrective, therapeutic, or traumatic reasons.
Subject has been treated with radiation to the face.
Subject was diagnosed with any major infectious processes, primary or secondary malignancies involving the face that have been active or required treatment in the past six (6) months.
Subject currently meets criteria as defined by the Diagnostic and Statistical Manual of Mental Disorders (DSM-IV-TR) for an active major depressive episode or for active significant psychiatric disorders, including dementia, uncontrolled general anxiety disorder, psychotic disorders or uncontrolled bipolar disorder. Exclusion criteria include: suicidality, active psychosis, untreated severe depression and/ or anxiety disorder, litigation, addiction, homicidal ideation, significant Axis II disorders and untreated sleep disorder.
Subject currently has clinically significant drug or alcohol abuse as defined by DSM-IV-TR or is unable to refrain from substance abuse throughout the study.
Subject is currently participating or has participated in the last month in another clinical study in which the subject has, is, or will be exposed to an investigational or non-investigational drug or device.
Subject is felt to be at risk of non-compliance (e.g. for completing the diary or maintaining a stable headache medicine regimen) in the investigator's opinion.
Subject is woman of childbearing age who is pregnant, nursing, or not using contraception.
Subject has had previous radio-frequency ablation of the SPG.
Subject has had blocks of the SPG in last 3 months.
Subject has undergone botulinum toxin injections of the head and/or neck in the last 3 months.
Subject has an implantable stimulator or any implanted devices in the head and/or neck.
Subject has H/O bleeding disorders or coagulopathy, or is on anticoagulation, antiplatelet, or GP IIb IIIa inhibitor medication.
Subject has H/O malignancy or any other condition that requires MRI monitoring.
Subject has or requires pacemaker/defibrillator.
Subject is not suitable for the study, in the judgment of the Investigators, due to social, co-morbid psychological, and/or medical considerations.
Subject has H/O stroke, cardiovascular disease, and/or epilepsy.
Subject was on antipsychotic or antidepressant medications (except for migraine prevention) in the past 3 months prior to the study.
Subject is allergic or has shown hypersensitivity to materials of the Medtronic components which come in contact with the body.
Primary purpose
Allocation
Interventional model
Masking
3 participants in 1 patient group