A team of researchers at Rice University and Queen Elizabeth Central Hospital (QECH) are working to develop a low-cost temperature sensor that can continuously monitor an infant's temperature (NTM). This robust, low-cost device will allow for the individualized monitoring of each infant with alerts for hypo and hyperthermia. A reusable band placed around the infant's abdomen to hold the temperature sensor will eliminate disposable components. This study will assess the accuracy of this novel device against a gold standard (Philips Intellivue patient monitor) and up to two existing devices (Bempu and Thermospot).
Full description
This study will assess the accuracy of this novel device against a commercially available patient monitor (Philips Intellivue MP30). Results will be compared to that of two existing devices designed for low resource settings: Bempu, a temperature monitor placed around an infant's wrist; and Thermospot, a sticker that changes color with temperature.
This study will enroll up to 150 infants at Queen Elizabeth Central Hospital. Up to 75 subjects will be collected from the neonatal ward and 75 subjects from the Kangaroo Mother Care (KMC) ward. This sample size will ensure that the investigators are able to collect sufficient data from infants with a range of gestational ages, weights, temperatures, and treatment locations (ie open cot, radiant warmer). The data from this study can be used to calculate the sample size needed in a larger study to evaluate changes in outcomes and nurse behavior related to the different temperature monitors (NTM, Bempu, Thermospot).
During the trial, the following steps will be taken:
A trained study nurse will assess the subject for clinical complications before attaching the temperature monitoring devices.
The goal of this study is to confirm the accuracy of these temperature monitors. Target accuracy for NTM device is +/- 0.5 C. In order to confirm this temperature accuracy in the settings of both the neonatal ward and the KMC ward, 75 subjects are needed in each setting. The international standards for clinical thermometers, standard number 80601-2-56, requires that 75 subjects from each patient population be tested to demonstrate the clinical accuracy of the thermometer. The clinical accuracy is reported using a measure of bias as well as the limits of agreement.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Primary purpose
Allocation
Interventional model
Masking
150 participants in 1 patient group
Loading...
Central trial contact
Rebecca Elias, MPH
Data sourced from clinicaltrials.gov
Clinical trials
Research sites
Resources
Legal