Evaluation of Ventricular Pacing Suppression Algorithms in Dual Chamber Pacemaker (LEADER)

K

Keimyung University

Status

Completed

Conditions

Heart Failure
Pacemaker
Sick Sinus Syndrome

Treatments

Device: Pacemaker with VpS® algorithm on
Device: Pacemaker with IRSplus algorithm on
Device: Pacemaker with Fixed long AV delay

Study type

Interventional

Funder types

Other
Industry

Identifiers

NCT03843242
2018-08-022

Details and patient eligibility

About

The recent study using IRSplus and VpS algorithm from Biotronik pacemaker showed the significant reduction in ventricular pacing to less than 3%.

The purpose of this study is to evaluate the efficacy of IRSplus and VpS algorithm in reducing ventricular pacing compared with conventional DDD pacing with a fixed AV delay.

Full description

The adverse cardiac outcomes due to right ventricular apical pacing with dual chamber pacemakers have been widely observed such as ventricular dyssynchrony resulting in reduced left ventricular function, increased risk of heart failure and atrial fibrillation. To minimize the ventricular pacing, manufacturers of pacemaker have made an effort to develop special algorithms designed to deliver right ventricular pacing only in case of demonstrated persistent long PR interval or repetitively lacking intrinsic ventricular activation.

Currently, there are three methods and algorithms are available with a pacemaker from the Biotronik SE & Co. KG to minimize right ventricular pacing as follows; DDD mode with fixed longer atrioventricular (AV) delay than intrinsic conduction time; Intrinsic rhythm support (IRSplus); Ventricular pacing suppression (VpS).

In the real world, it is the most common practice to program the DDD(R) mode with fixed long AV delay because of physician's concerns about the possible failure of an algorithm or long pause being resulted from 2 consecutive loss of AV conduction by the algorithm. However, the main disadvantage of a fixed with long AV delay is that the prolonged total atrial refractory period (TARP) results in changes of the upper rate behavior (i.e., pseudo-Wenckebach AV block and subsequently 2:1 block at lower atrial tracking rates). The possibility of pacemaker-mediated tachycardia is getting high if the post-ventricular atrial refractory period (PVARP) is shortened to compensate. Furthermore, there is a high chance to have fusion/pseudo-fusion of ventricular pacing in a fixed long AV delay because the AV conduction is dynamic according to the heart rate.

The recent systematic review showed that there are no significant differences between the pacing modes for mortality, heart failure, stroke, and atrial fibrillation (AF) in patients with sinus node dysfunction (SND) without AV block. However, the dual chamber pacemaker is still recommended in patients with SND due to lack of tools to identify patients at high risk of developing the complete AV block. The meta-analysis about the effect of the reduction in unnecessary ventricular pacing using a sophisticated algorithm in patients with SND showed there are no benefits in clinical outcomes compared with conventional DDD mode. But the percentage of ventricular pacing in ventricular pacing reduction modality group was not negligible between 1~11.5%. The result might be changed if we have data with a more significant reduction in ventricular pacing using difference algorithm. The recent study using IRSplus and VpS algorithm from Biotronik pacemaker showed the significant reduction in ventricular pacing to less than 3%.

The purpose of this study is to evaluate the efficacy of IRSplus and VpS algorithm compared with conventional DDD pacing with a fixed AV delay.

Enrollment

146 patients

Sex

All

Ages

20+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • Sick sinus syndrome
  • No evidence of 2nd and 3rd degree AV block
  • Provide written informed consent
  • Age ≥ 20 years old

Exclusion criteria

  • 2nd and 3rd degree AV block
  • History of AF
  • patients with older version of pacemaker
  • Life expectancy ≤ one year
  • Pregnant or lactating women

Trial design

Primary purpose

Diagnostic

Allocation

Randomized

Interventional model

Crossover Assignment

Masking

Single Blind

146 participants in 3 patient groups

Pacemaker with Fixed long AV delay
Experimental group
Description:
1. Patients who meet the inclusion criteria and is implanted with a Biotronik Enitra 8 DR-T pacemaker are eligible. 2. The pacemaker was programmed with a long and fixed atrioventricular interval for the first 3 months. 3. Definition of fixed AV delay (than intrinsic AV conduction) • If P-wave exists: intrinsic AV conduction time = As \~ Vs interval in the marker channel sensed AV delay = intrinsic AV conduction time + 20 msec paced AV delay = sensed AV delay + 30 msec • If no P-wave exits: intrinsic AV conduction time = Ap \~ Vs interval in the marker channel paced AV delay = intrinsic AV conduction time + 20 msec sensed AV delay = paced AV delay - 30 msec • If the intrinsic AV conduction time is ≥ 300ms, make paced/sensed AV delay 350/320 msec
Treatment:
Device: Pacemaker with Fixed long AV delay
Pacemaker with VpS® algorithm on
Experimental group
Description:
Vp Suppression ON algorithm: This feature promotes the intrinsic AV conduction by only pacing the ventricle when intrinsic conduction becomes unstable or disappears. Depending on the presence or absence of AV conduction, the feature is implemented either in the ventricular pacing suppression state ADI(R), which promotes the intrinsic conduction, or in the DDD(R) ventricular pacing state Vp DDD(R), which provides ventricular pacing. Automatic switching capabilities between those two states promotethe intrinsic conduction as much as possible without harming the patient. Scheduled Vs searching tests look for intrinsic conduction using an extended AV delay of 450ms.
Treatment:
Device: Pacemaker with Fixed long AV delay
Device: Pacemaker with VpS® algorithm on
Pacemaker with IRSplus algorithm on
Experimental group
Description:
IRS plus algorithm: This algorithm incorporates two different functions: the first is scan hysteresis, which better enables the heart to pace on its own by periodically extending the search time for its natural pacing stimulus (the intrinsic AV conduction) over six consecutive atrial cycles. The second is the repetitive hysteresis, which recognizes when the heart is not pacing on its own (a consistent loss of intrinsic AV conduction lasting for six consecutive atrial cycles) and switches the mode of the device from extended to basic atrioventricular (AV) delay.
Treatment:
Device: Pacemaker with Fixed long AV delay
Device: Pacemaker with IRSplus algorithm on