Status
Conditions
Treatments
About
This is a pivotal, non-randomized, single arm, multi-center, prospective, non-significant risk study to evaluate the FreeStyle Libre Flash Glucose Monitoring System.
Full description
Up to 60 subjects will be enrolled at up to six clinical research sites in the United States. Subjects will wear two Sensors of Three unique sensor lots. Each Sensor will have a paired Reader that will be given to the subject. All Readers will be masked during the study (i.e. subjects will not be able to view glucose results obtained from the Sensor on the Reader screen).
Subjects will be asked to perform at least 8 capillary Blood Glucose (BG) tests per day using the primary Reader. Interstitial glucose readings from each Sensor will be obtained with the corresponding Readers immediately following each BG test. Subjects will follow the same BG testing schedule during in-clinic visits. Subjects will be instructed to report any problems with the device.
Subjects will make seven (7) scheduled visits to the clinical study site, including the Enrollment/Screening Visit (Visit 1). Subjects will have four (4) in-clinic visits during which intravenous blood draws and YSI reference testing will occur. Each in-clinic visit is anticipated to last approximately 10 hours.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
50 participants in 1 patient group