This study is a non-randomized, single-arm, multi-center study that is designed to evaluate the safety and effectiveness of the FreeStyle Libre Flash Glucose Monitoring Systems in pediatric populations.
Full description
Up to 85 subjects will be enrolled at up to four (4) clinical research sites in the United States. Subjects will wear two Sensors. Each Sensor will have a paired Reader that will be given to the subject. All Readers will be masked during the study (i.e. subjects will not be able to view glucose results obtained from the Sensor on the Reader screen). Subjects will be asked to perform at least 4 capillary Blood Glucose (BG) tests per day using the primary Reader. Interstitial glucose readings from each Sensor will be obtained with the corresponding Readers immediately following each BG test. Subjects will be instructed to report any problems with the device. Subjects will make four (4) to five (5) scheduled visits to the clinical study site, including the Enrollment/Screening Visit (Visit 1). Based on the subjects weight, subjects will have up to two (2) in-clinic visits during which intravenous blood draws and YSI reference testing will occur.