This study will test how accurate a blood pressure cuff is during certain types of surgery.
During some abdominal surgeries done with small tools and a camera (called laparoscopic surgery), doctors place patients in a head-down position called the Trendelenburg position. In this position, it may be harder for the blood pressure cuff on the arm to give correct readings. This study will compare the cuff's readings (non-invasive) with readings from a thin tube placed inside an artery (invasive), which is more accurate but also more risky and uncomfortable.
The researchers will use a method called error grid analysis to check how closely the cuff readings match the invasive ones. They hope to learn if the blood pressure cuff is accurate enough to be used safely in this kind of surgery. If it is, doctors may not need to use the invasive method as often.
People 18 years or older who are already going to have invasive blood pressure monitoring as part of their planned surgery can join the study. The study will collect blood pressure measurements about every 20 minutes during surgery, as well as basic information like age, weight, and what medicines are given.
Full description
This is a prospective observational study designed to evaluate the clinical accuracy and reliability of non-invasive blood pressure (NIBP) monitoring in adult patients undergoing laparoscopic surgery in the Trendelenburg position. The accuracy of NIBP will be assessed by comparing it to invasive arterial blood pressure (IABP), which is considered the reference standard. The comparison will be performed using error grid analysis, a method that not only quantifies measurement differences but also categorizes the clinical risk of those discrepancies.
The Trendelenburg position, characterized by a head-down tilt of the patient, is frequently employed during gynecological and urological laparoscopic procedures to enhance surgical exposure. However, this positioning alters hemodynamic physiology and may influence peripheral blood pressure measurements. Previous evaluations of NIBP in this context have primarily relied on Bland-Altman analysis, which measures statistical agreement but does not assess the potential clinical consequences of disagreement between methods.
Error grid analysis overcomes this limitation by mapping paired NIBP and IABP values into predefined risk zones (A-E), based on their potential to affect clinical decision-making. Zone A reflects acceptable agreement with no risk of incorrect treatment, whereas Zones B through E indicate increasing levels of clinical risk.
In this study, NIBP and IABP will be measured simultaneously every 20 minutes throughout the procedure. The proportion of measurements falling within each error grid zone will be calculated. If fewer than 90% of the paired measurements fall within Zone A, NIBP will be considered insufficiently reliable for use in this surgical context without concurrent invasive monitoring.
A pilot study of five patients was conducted to estimate the intracluster correlation coefficient and design effect, which informed the final sample size calculation. Based on these preliminary data, a total of 100 patients is estimated to provide adequate statistical power to evaluate the primary outcome with 95% confidence and a 5% margin of error.
The study will also document intraoperative variables that may affect blood pressure measurement accuracy, including the use of vasoactive infusions. These data may provide insight into subgroups where NIBP reliability is particularly limited or acceptable.
The findings of this study may inform future monitoring protocols and could reduce unnecessary use of invasive lines in appropriately selected patients, improving patient safety and comfort while maintaining clinical accuracy.