The study objectives are to assess the safety and efficacy of the CoreValve Evolut R transcatheter aortic valve replacement (TAVR) system in patients with severe symptomatic aortic stenosis are considered at high or extreme risk for surgical aortic valve replacement.
Full description
This objective will be accomplished by a prospective, single arm, historical controlled, multi-site study involving a minimum of 150 implanted subjects with no more than 250 implanted subjects at up to 25 study sites in the United States. Procedural and 30 day safety and efficacy results from this study will be compared to appropriate historical control data for the Medtronic CoreValve System. Subjects will be followed up to 5 years following implantation.
The enrollment phase of the study is estimated to take approximately 6-9 months. As each implanted subject is to be followed up to 5 years, the estimated study duration is approximately 66-69 months, excluding the time required for preparing the final report.
Enrollment
Sex
Volunteers
Inclusion and exclusion criteria
Inclusion Criteria - Severe aortic stenosis, defined as aortic valve area of < 1.0 cm2 (or aortic valve area index of < 0.6 cm2/m2) by the continuity equation, AND mean gradient > 40 mmHg or maximal aortic valve velocity > 4.0 m/sec by resting echocardiogram.
Subjects with low-flow/low gradient severe aortic stenosis can be included, provided low-dose dobutamine or exercise stress echocardiography demonstrates a mean gradient of >40 mmHg or a maximal aortic valve velocity of >4.0 m/sec, AND aortic valve area of <1.0cm2 (or aortic valve area index of <0.6 cm2/m2).
Exclusion Criteria
Anatomical exclusion criteria:
For transfemoral or transaxillary (subclavian) acess:
Primary purpose
Allocation
Interventional model
Masking
241 participants in 1 patient group