Status
Conditions
Treatments
About
The goal of this proposed study is to compare use of a PID (Proportional-Integral-Derivative) controller versus an MPC (Model Predictive Control) controller algorithm in an artificial pancreas system, all other components and study design being equal. The study design, power calculation and endpoints were developed based on the results of an initial feasibility study (ClinicalTrials.gov Identifier: NCT01987206) that has already been completed.
Full description
This randomized crossover study consists of an evaluation of either type of control algorithm (MPC or PID) as a part of the Artificial Pancreas (AP) device during two periods of 27.5-hour closed-loop control in a minimally supervised setting (outpatient research area at the William Sansum Diabetes Center, Santa Barbara, CA) separated by a minimum of 5 days and a maximum of 2 weeks. The 27.5-hour period includes: 2 announced meals (dinner and breakfast of 65g and 50g CHO respectively) preceded with a dose of rapid-acting insulin equivalent to 100% bolus based on each subject's Insulin to Carbohydrate (I:C) ratio and 1 unannounced meal (lunch of 65g carbohydrates, same meal content as dinner); complete night from 12:00 am to 7:00 am. The goal is to demonstrate that the AP device is able to maintain the subject blood glucose within a safe range at all times.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Hematocrit < 30% or >55% A1C > 10% Abnormal liver or renal function (Transaminase >2 times the upper limit of normal, Creatinine> 1.5 mg/dL) Labs drawn at screening visit or within one month prior to screening (for other purposes) will suffice for enrollment purposes related to hematocrit
Primary purpose
Allocation
Interventional model
Masking
20 participants in 3 patient groups