Non-Invasive Blood Pressure Accuracy Study for Adult Subject Populations

General Electric (GE) logo

General Electric (GE)

Status

Enrolling

Conditions

Blood Pressure Measurement
Blood Pressure

Treatments

Device: NIBP Measurement

Study type

Interventional

Funder types

Industry

Identifiers

NCT07628296
SA-000259

Details and patient eligibility

About

The purpose of this study is to collect NIBP and invasive aortic pressure data from the clinical setting.

Full description

The purpose of the study is to collect clinical data for validation of the accuracy of non-invasive blood pressure (NIBP) measurement algorithms on the Non-invasive Blood Pressure (NIBP) Standard Parameter Module (SPM) per IEC 80601-2-30:2018, Medical electrical equipment-Part 2-30: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers standard, section 201.106. In addition, the study will be conducted per ISO 81060-2:2018+AMD2:2024 Non-invasive sphygmomanometers-Part 2: Clinical investigation of intermittent automated measurement type.

Enrollment

130 estimated patients

Sex

All

Ages

18+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  1. Age requirement: 18 and over.
  2. Have an upper limb (right OR left side) that fits the cuff size of the device (circumference ranging 17 cm to 40 cm) OR have a thigh (right OR left side) that fits a cuff size of the device (circumference ranging 38 to 50 cm).
  3. Have an arterial line as a standard of care, which measures blood pressure at the aortic root.
  4. Able to provide written informed consent.
  5. Expected to remain in the unit for the duration of the study procedure.

Exclusion criteria

  1. Have previously participated in this study (no subject may participate more than once).
  2. Pregnant and lactating women per site standard of care.
  3. Subjects with peripheral arterial diseases (causing left/right arm difference)
  4. Hemodynamically unstable subjects at the discretion of the investigator.
  5. Exhibited injuries, deformities, intravenous lines, or other abnormalities that, in the opinion of the Investigator, may prevent proper cuff application or functioning.
  6. Are suffering from infection(s) or immunocompromised patients that require isolation.
  7. PI or designee decision due to subject health conditions

Trial design

Primary purpose

Diagnostic

Allocation

Randomized

Interventional model

Single Group Assignment

Masking

None (Open label)

130 participants in 3 patient groups

Small Adult vs Adult Cuff (Upper Arm, Overlap Range 23-25 cm)
Experimental group
Description:
Subjects with an upper-arm circumference within 23-25 cm, where both Small Adult (SA) and Adult (A) cuffs are suitable, are randomized in a 1:1 ratio to receive blood pressure measurements using either: * Small Adult cuff, or * Adult cuff Subjects with upper-arm circumferences in non-overlapping SA-only ranges are assigned the Small Adult cuff without randomization and are not included in this arm.
Treatment:
Device: NIBP Measurement
Adult vs Large Adult Cuff (Upper Arm, Overlap Range 31-33 cm)
Experimental group
Description:
Subjects with an upper-arm circumference within 31-33 cm, where both Adult (A) and Large Adult (LA) cuffs are suitable, are randomized in a 1:1 ratio to receive blood pressure measurements using either: Adult cuff, or Large Adult cuff Subjects with measurements in non-overlapping A-only or LA-only ranges are assigned the appropriate cuff without randomization and are not included in this arm.
Treatment:
Device: NIBP Measurement
Thigh Cuff (Fallback/Non-Randomized)
Experimental group
Description:
Subjects whose arm circumference is either: * Outside the allowable upper-arm range (17-40 cm), or * Eligible but excluded due to fulfillment of upper-arm cuff quotas will undergo thigh circumference measurement. Those with a thigh circumference within 38-50 cm will receive blood pressure measurements using a thigh cuff. Subjects outside the thigh circumference range are excluded prior to any blood pressure measurements.
Treatment:
Device: NIBP Measurement