Noninvasive Monitoring of Vital Signs in Neonates

Clalit Health Services logo

Clalit Health Services

Status and phase

Unknown
Phase 2

Conditions

Admission to the NICU and Need for Physiological Monitoring.

Treatments

Device: The new sensor is not on the market. We have named it Elfi-sensor.

Study type

Interventional

Funder types

Other

Identifiers

NCT01707589
Pediatric Elfi-Monitor

Details and patient eligibility

About

the investigators are assessing a non-invasive monitor based on dynamic light scattering that measures pulse rate and respiration rate. The monitor also will be equipped with a temperature sensor and a motion sensor. This monitor will allow multi-parametric measurements in neonates in a simple and noninvasive manner.

Full description

Monitoring vital signs in neonates is important. This currently is done with multiple standard clinical monitors. the investigators are developing a very small (coin-sized) sensor that can measure pulse rate, respiratory rate, temperature and body motion. the investigators propose to test our sensor against standard monitoring techniques in neonates in a neonatal ICU (NICU).

The monitor, called ELFI-monitor, is based on dynamic light scattering. A low-power laser beams light into the skin, and the red blood cells in the underlying skin return the light which is recorded in a light sensor. The movement of the red blood cells is affected by the rhythmic cardiac contractility. in this manner, heart rate and rhythm can be assessed. The ELFI monitor also contains an embedded temperature probe and a miniature 3-D motion sensor, allowing capture of additional parameters.

Methods:

the investigators plan to study 100 neonates admitted to the NICU of Meir Hospital Center. These infants are routinely monitored with standard equipment including ECG, oximeter, temperature, and respiratory monitors. the investigators will simultaneously monitor these children with the ELFI monitor, recording pulse rate, cardiac rhythm, respiratory rate, skin temperature and body motion. Recordings will be conducted for a one-hour period on each subject. Data recordings from the ELFI sensor will be compared to the readouts of the conventional monitors used on the patient. It is important to emphasize that the data recordings of the ELFI sensor will not affect clinical decisions in any manner, but will merely be recorded for comparison with conventional sensors.

Enrollment

100 estimated patients

Sex

All

Ages

1 day to 6 months old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • All neonates admitted the Meir Hospital center NICU
  • Parents provide informed consent

Exclusion criteria

  • Neonates with skin disease

Trial design

Primary purpose

Diagnostic

Allocation

Non-Randomized

Interventional model

Single Group Assignment

Masking

None (Open label)

100 participants in 1 patient group

Neonates admitted to the NICU
Experimental group
Description:
Neonates admitted to the NICU for a variety of medical reasons will be the cohort group. Those whose parents give informed consent will compose the subjects of the study
Treatment:
Device: The new sensor is not on the market. We have named it Elfi-sensor.