Oximeters and Skin Pigment: Impact of Skin Pigmentation on Accuracy of Pulse Oximetry

University of Cincinnati logo

University of Cincinnati

Status

Completed

Conditions

Hyperoxia
Hypoxemia
Pigment Skin

Treatments

Device: SpO2 and SaO2 comparison

Study type

Observational

Funder types

Other
Other U.S. Federal agency

Identifiers

NCT05690269
HRP-503

Details and patient eligibility

About

Finger pulse oximetry SpO2 is widely used to estimate arterial oxygen saturation SaO2. Current Spo2 targets do not take into consideration the many variables that affect that correlation in particular skin pigmentation.

This study aims to evaluate different FDA approved pulse oximeters (Nonin co-pilot, massimo Radical 7, Philips-standard of care monitor, innovo premium iP900BP, nellcor PM1000N, Nano100) with SaO2 reference values obtained by an arterial blood gas in subjects with different skin pigments measured by a skin color scale as well as self-identification of race.

Full description

The oximeters that will be evaluated are nonin co-pilot, massimo Radical 7, Philips (standard care), Innovo Premium iP900BP, nellcor PM1000N and USA medical supply Nano100. The sensors used will be reusable digital sensors for all oximeters. Oximeters will have been placed on patients' fingers (3 fingers on each hand) during an initial phase of clinical stability. Due to the impact of the fingers used for SpO2 measurement and bias that might be caused by right hand versus left hand, oximeters will be randomized to different fingers and hands for each patient. Measurements will be made when the SpO2 with the usual oximeter is <97% as recommended and in the absence of changes of the respiratory parameters in the last 10 minutes. A set of measures that will include all the oximeters tested will be used. Once the oximeters are installed, at least 2 minutes will run out to let the oximeters reach steady state. Then when a standard of care blood gas is obtained, measurements will be taken simultaneously to the drawing of a standard care arterial blood gas by a clinician in the ICU. While the clinician takes the arterial blood gas, a picture will be simultaneously taken of the SpO2 readings shown by each oximeter, in addition to taking vital signs, O2 flow or FiO2.

Enrollment

232 patients

Sex

All

Ages

18+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • ≥ 18 years old
  • Patients admitted to the Intensive Care Unit
  • Patient has arterial line

Exclusion criteria

  • No signal with the oximeter
  • Missing digits preventing application of all probes simultaneously
  • No requirement for arterial blood gases
  • Pigmented nails, fake nails, or nail polish
  • Methemoglobinemia
  • Patient in isolation (multi-resistant bacteria, C-Difficile, COVID-19)