Performance Evaluation of Blood Glucose Monitoring Systems (Polaris)

I

IKFE Institute for Clinical Research and Development

Status and phase

Completed
Phase 4

Conditions

Diabetes Mellitus

Treatments

Other: Sugar infusion
Drug: insulin infusion

Study type

Interventional

Funder types

Other
Industry

Identifiers

NCT01504620
BGSTA_C_05445
SAN-BGM-002 Precision (Other Identifier)
SAN-HCT-002 Helios (Other Identifier)

Details and patient eligibility

About

The goal of this single-center, comparative, open label, in-vitro diagnostic device performance evaluation was to investigate the performance of two sanofi-aventis BGMSs (BGStar® and iBGStar®) with regard to system accuracy (Polaris), precision and Hct influence (Helios).

Full description

The main study (Polaris) investigated blood glucose measuring accuracy and intra-assay precision while the substudy (Helios) investigated potential haematocrit interference of sanofi-aventis BGMSs BGStar® and iBGStar® and partly of competitor BGMSs according to ISO 15197 [1] and TNO 2001 Quality Guideline [2], respectively.

The primary objective of the system accuracy evaluation is to identify the number of blood glucose readings of both sanofi-aventis devices within the ISO 15197 area of acceptable accuracy of ± 20% [DEV%] for blood samples with glucose concentration > 75 mg/dL (4.2 mmol/L) and within the ISO 15197 area of acceptable accuracy of ± 15 mg/dL (± 0.83 mmol/L) for blood samples with glucose concentration < 75 mg/dL (4.2 mmol/L) as compared to the StatStrip® Connectivity reference method. The primary objective of the haematocrit interference evaluation is to demonstrate that the BGStar® and iBGStar® devices meet the Hct interference requirements according to TNO 2001 Quality Guideline when used to measure BG of venous whole blood as compared to a venous plasma reference method (YSI 2300 STAT Plus™) at Hct levels between 35 - 55% in terms of:(1) a mean deviation not exceeding ± 10% [DEV%] for hyperglycaemic glucose concentrations (150 - 180, 250 - 280, 320 - 350 mg/dL) and (2)a mean deviation not exceeding ± 18 mg/dL (1.0 mmol/L) [DEVtotal] for hypo- and normoglycaemic glucose values (50 - 60, 100 - 120 mg/dL) compared to the reference method at Hct levels between 35 - 55%.

Enrollment

106 patients

Sex

All

Ages

18 to 75 years old

Volunteers

Accepts Healthy Volunteers

Inclusion criteria

  • healthy subjects or patients with type 1 or type 2 diabetes
  • screening glucose levels fall into required glucose ranges
  • screening hematocrit values between 36 % to 55 %

Exclusion criteria

  • history of hypotension during blood draws
  • intake of drugs known to interfere with blood glucose readings
  • biochemical safety parameters outside of reference ranges
  • Hb < 11 g/dL
  • lack of compliance
  • history of frequent hypoglycemia

Trial design

Primary purpose

Diagnostic

Allocation

Non-Randomized

Interventional model

Single Group Assignment

Masking

None (Open label)

106 participants in 2 patient groups

Sugar infusion
Other group
Description:
Diagnostic assessment of blood glucose by means of different devices
Treatment:
Other: Sugar infusion
insulin infusion
Experimental group
Description:
infusion of insulin to achieve low glucose levels
Treatment:
Drug: insulin infusion