ClinicalTrials.Veeva

Menu

Portable Artificial Pancreas Applied for Youth and Adolescents (PAPAYA 1)

I

Inreda Diabetic

Status

Completed

Conditions

Diabetes Mellitus, Type 1

Treatments

Device: Open-loop control
Device: Bi-hormonal closed-loop control

Study type

Interventional

Funder types

Industry
Other

Identifiers

NCT05543850
NL79829.000.22

Details and patient eligibility

About

The purpose of this study is to determine the performance of a bi-hormonal reactive closed-loop system in adolscents with type 1 diabetes mellitus. The percentage of time spent in the target range (3.9-10.0 mmol/L) is the main outcome and will be compared between the bi-hormonal closed-loop system and the current treatment of the patients. Also, safety parameters, pharmacodynamics and patient reported outcomes (expectations, trust and treatment satisfaction scores) are compared.

This study is a monocenter, randomized, cross-over trial with 20 subjects. The subjects will be randomized to receive either the open-loop therapy or the closed-loop therapy for the first two-week study period and switch to the alternate treatment with at least two weeks in between.

Full description

Background of the study:

Inreda Diabetic B.V. (Goor, The Netherlands) developed a bi-hormonal reactive closed loop system to automate glucose regulation (artificial pancreas; AP) in patients with diabetes mellitus type 1. In previous studies, we tested the performance and safety of this bi-hormonal closed-loop system in an outpatient setting in patients with type 1 diabetes mellitus (T1DM) for a duration up to 2 weeks, leading to CE-marking of the system. The device is currently intended for insulin-dependent adult patients with diabetes. However, there is great need to improve glycaemic control in youth and adolescents with diabetes.

Objective of the study:

The main objective of this study is to determine the performance of the closed-loop system in adolescents with T1DM. Secondary objectives include: to assess the safety of the closed-loop system; to determine the time that the closed-loop algorithm is active; and to assess expectations and treatment satisfaction of the closed-loop system.

Study design:

This study is a monocenter randomized cross-over trial.

Study population:

The study population will comprise 20 patients with T1DM, between 12 and 18 years old, not having impaired awareness of hypoglycaemia, and who are treated with insulin therapy for at least 6 months. Patients receiving treatment with multiple daily injections (MDI) or continuous subcutaneous insulin injections (CSII), and in addition glucose monitoring using self-monitored blood glucose (SMBG), flash glucose monitoring (FGM) or continuous glucose monitoring (CGM) are included into the study.

Intervention:

The intervention includes 2 weeks of closed-loop control with the artificial pancreas (AP) of Inreda Diabetic. The device uses 2 subcutaneous glucose sensors, 2 subcutaneous infusion sets, and incorporates 2 pumps and a reactive closed-loop algorithm. During 4-6 days before the intervention the patients receive training on the use of the closed-loop system and will start using the device under close supervision. The control arm (open loop treatment) consists of the patient's standard therapy at home for 2 weeks.

Main study parameters/endpoints:

Main study parameter is the percentage of time spent in the target range (3.9-10.0 mmol/L), which will be compared between the open and closed-loop periods for each patient.

Enrollment

20 patients

Sex

All

Ages

12 to 18 years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • Diagnosed with type 1 diabetes mellitus
  • Twelve to eighteen years old;
  • Treated with insulin therapy for at least 6 months;
  • Willing and able to sign informed consent or to assent to study participation.

Exclusion criteria

  • Impaired awareness of hypoglycaemia (score ≥ 4) according to Gold and/or Clarke questionnaire[6,7];
  • BMI ≥ 35 kg/m2;
  • Pregnancy and/or breastfeeding;
  • HbA1c > 97 mmol/mol (11.0%);
  • Use of acetaminophen (paracetamol) during the open loop or closed-loop period,as this may influence the sensor glucose measurements;
  • Limited ability to see, hear or feel the alarm signals of the closed-loop system;
  • Unwillingness to act in response to the alarm signals;
  • Living alone during the night during the closed-loop period (the patient may ask someone to stay over temporarily);
  • Expected poor internet connectivity regarding 24/7 tele monitoring;
  • Any condition that the local investigator feels would interfere with trial participation or evaluation of the results.

Trial design

Primary purpose

Treatment

Allocation

Randomized

Interventional model

Crossover Assignment

Masking

None (Open label)

20 participants in 2 patient groups

Open loop control
Active Comparator group
Description:
In this arm, the patients will use their usual diabetes therapy. Additionally, patients will wear a blinded DexcomG6 for data collection during the open loop period.
Treatment:
Device: Open-loop control
Bi-hormonal closed-loop control
Experimental group
Description:
In this arm, treatment consists of the bi-hormonal closed-loop system. A short acting insulin analogue (Humalog, 3 ml pre-filled cartridge, Eli Lilly) and glucagon (Glucagen, Novo Nordisk; 3.15 ml cartridge Accu-Chek Spirit, Roche) will be administered according to the closed-loop algorithm. Additionally, patients will wear a blinded DexcomG6 for data collection during the closed loop period.
Treatment:
Device: Bi-hormonal closed-loop control

Trial contacts and locations

1

Loading...

Central trial contact

Helga Blauw, PhD

Data sourced from clinicaltrials.gov