ClinicalTrials.Veeva

Menu

Project VerioVue Enhancements - Neonate Study

L

LifeScan

Status

Terminated

Conditions

Diabetes Mellitus

Treatments

Device: VerioVue (Enhancements) blood glucose monitoring system

Study type

Observational

Funder types

Industry

Identifiers

NCT06052371
FILE-PROT-005443

Details and patient eligibility

About

The goal of this performance evaluation is to compare the blood glucose results obtained using the VerioVue Blood Glucose Monitoring system (BGMS) with those obtained from a validated comparator method (iSTAT 1 Analyser) using blood obtained from neonates. The main question it aims to answer is:

•How accurate is the VerioVue BGMS when compared to a product that has already been confirmed as accurate (iSTAT 1 Analyser) when hospital staff test blood taken from neonates on these instruments? Participants (neonates) will have a small amount of blood taken from a heel prick (performed by a HCP for medical purposes) or from an existing arterial line.

Full description

UK NHS Hospital sites (between 1 and 3 sites) will be used to collect a minimum of 200 evaluable blood samples from neonates meeting the inclusion criteria. Consent will be taken from the neonates parent prior to participation in the study. They will consent to one or two blood samples to be taken from their child. If the participant meets the enrolment criteria then demographics and prescription medication (including dietary supplements) information will be collected. A small volume of blood (max 150µl) will be collected from an existing arterial line or from a heel prick that is being performed for medical purposes. This blood will then be applied to six investigational One Touch VerioVue meters (with expanded haematocrit range of 20-65%) fitted with OneTouch Verio test strips and the blood glucose meter readings recorded. Three different lots of test strips will used in rotation across the study. Within 5 minutes of the last meter test, NHS study staff will use blood from the same sample and perform a test on the iSTAT 1 Analyser to obtain blood glucose, haematocrit and oxygen levels of the sample. No treatment decisions will be made based on the blood glucose results obtained for this study. The study documentation will be reviewed by LifeScan monitors for completeness and accuracy and the results entered into a validated database. The data will be extracted and statistically analysed to determine the accuracy of the blood glucose results obtained using the OneTouch VerioVue BGMS when compared to the blood glucose results obtained on the iSTAT 1 Analyser.

Enrollment

99 patients

Sex

All

Ages

Under 28 days old

Volunteers

Accepts Healthy Volunteers

Inclusion criteria

  • Age - Newborn babies, 28 days old or less with a corrected gestational age of at least 34 weeks and a weight of at least 1700g or more at the time of participation.
  • Informed Consent - Study participants parents must read the Parental Participant Information Sheet and sign the Parental Informed Consent Form
  • Study participants parent agrees to provide information related to demographics, prescription medication, dietary supplements, and results from other physician ordered testing e.g., Bilirubin (total, conjugated, and unconjugated) where available.
  • Study participants parent agrees to allow study staff access to medical records where necessary.
  • Study participants parents agree to all aspects of the study process, including where applicable arterial blood draw from an existing arterial line or a heel prick performed by an HCP for medical purposes.

Exclusion criteria

  • Age- Newborn more than 28 days old.
  • Study participants with a gestational age of less than 34 weeks at the time of participation.
  • Study participants with a gestational weight of less than 1700g
  • Current positive test result for Covid-19.

Trial design

99 participants in 1 patient group

Neonates
Description:
Newborn babies who are 28 days old or less with a corrected gestational age of at least 34 weeks and a weight of at least 1700g or more at the time of participation. They will be recruited from hospital neonatal units or wards.
Treatment:
Device: VerioVue (Enhancements) blood glucose monitoring system

Trial contacts and locations

3

Loading...

Central trial contact

Ayaz Alam; Barry Irvine

Data sourced from clinicaltrials.gov