Aerosol delivery during mechanical ventilation has long been a long debated topic. As evidence-based knowledge about the delivery of aerosol to the lungs of mechanically ventilated patients increases, one piece of the puzzle has remained unexplored; measurement of clinically relevant outcomes. The primary aim of this research is to compare clinical outcomes (ventilator-associated events (VAEs), length of stay (LOS) in intensive care unit (ICU), and total days on mechanical ventilation) when using a traditional jet nebulizer versus a newer generation vibrating mesh nebulizer during mechanical ventilation. The secondary aim of this research is to identify source of bacteria by obtaining cultures of each nebulizer and ventilator circuit and plating them for colony growth and identification.
Full description
This pilot study was designed as a randomized, controlled trial comparing clinical outcomes and contamination rates in jet nebulizers versus vibrating mesh nebulizers. The sample size will be one of consecutive convenience over an entire year (September 2013-September 2014) to include all seasons. The study will be performed at Rush University Medical Center (RUMC) in Chicago, IL.
Each mechanically ventilated patient within inclusion criteria, with a physician order for aerosol treatment, will be randomized to either the jet nebulizer or mesh nebulizer group using SPSS computer software. The respiratory care staff will place the nebulization device in the ventilator circuit per hospital protocol. Both devices, jet and mesh nebulizers, are currently standard practice; therefore hospital protocol will be followed regarding placement and administration of ordered aerosol treatment.
The study subject will remain on the device to which they were randomized for the duration of their hospital stay to obtain clinical outcome data. Study staff to be notified of patient extubation from mechanical ventilation, discharge from intensive care unit, or expiration; this will conclude participant involvement in the study. Retrospective clinical outcome data will be obtained from the subject's electronic medical chart at conclusion of study.
Retrospective data to include:
LOS in ICU (days)
LOS on mechanical ventilator (days)
Drug administration information
Drug and dose ordered
Number of treatments delivered
Treatment type
Ventilator-Associated Tracheobronchitis (# VATs)16
a. Fever, increased volume and purulence of secretions, a positive culture (quantitative or semi-quantitative) of a respiratory sample (tracheal aspirates or bronchoscopic specimens), and the absence of a new or an evolving pulmonary infiltrate in the chest x-ray in a patient on mechanical ventilation >48 hours
Ventilator-Associated Events in Adults15
Ventilator-Associated Condition (# VACs)
Infection-related Ventilator-Associated Complication (# IVACs)
Possible Ventilator-Associated Pneumonia (# Possible VAPs)
Probably Ventilator-Associated Pneumonia (# Probable VAPs)
The Jet Nebulizer Protocol:
Physician order received for subject; randomization to jet nebulizer occurred.
Jet nebulizer, Misty Max 10™ (Carefusion, CA), placed into spring-loaded t-piece between the humidifier and the ventilator (approximately 15 cm from the gas outlet).
Aerosol treatment delivered per physician order at flow rates of 8-10 L/min.
Jet nebulizer to be replaced every 3 days per hospital protocol.
Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.
Cultures
Jet Nebulizer Culture (3 plates): Sample to be collected from jet nebulizer by running oxygen at a flow of 8-10 L/min through nebulizer (3 mL normal saline in the reservoir) while directing the nebulizer towards plate for 1 minute. Process to be repeated for 3 different plates. Plates to be labeled, have edges sealed with tape, be placed in biohazard bag and sent to microbiology lab for analysis.
Ventilator Circuit Culture: Material to be collected from rim of inspiratory ventilator circuit at wye by rubbing a sterile swab along circuit 3x in a circular motion. Swab to be sealed in test tube with lid and sent to lab for analysis.
Sputum samples: Data obtained per electronic medical chart. Sputum sample data to be analyzed in study when ordered by physician.
After obtaining cultures, the jet nebulizer will be replaced with new one and labeled per hospital protocol.
Further cultures are not necessary unless patient is identified by study staff to have VAE. (Appendix A)
Study staff to be notified of patient extubation from mechanical ventilation, discharge from intensive care unit or expiration.
The Vibrating Mesh Nebulizer protocol:
Physician order received for subject; randomization to vibrating mesh nebulizer occurred.
Aeroneb® Solo (Aerogen, Galway, Ireland) vibrating mesh nebulizer to be placed before the heater on the "dry side" of the heater water chamber on the inspiratory ventilator circuit.
Aerosol treatment delivered per physician order.
Mesh nebulizer to be replaced every 30 days per hospital protocol.
Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.
Cultures
Vibrating Mesh Nebulizer Culture (3 plates): Sample to be collected from vibrating mesh nebulizer while directing the nebulizer(3 mL normal saline in the reservoir) towards plate for 1 minute. Process to be repeated for 3 different plates. Plates to be labeled, have edges sealed with tape, be placed in biohazard bag and sent to microbiology lab for analysis.
Ventilator Circuit Culture: Material to be collected from rim of inspiratory ventilator circuit at wye by rubbing a sterile swab along circuit 3x in a circular motion. Swab to be sealed in test tube with lid and sent to lab for analysis.
Sputum samples: Data obtained per electronic medical chart. Sputum sample data to be analyzed in study when ordered by physician.
Vibrating mesh nebulizer replaced with new one and labeled per hospital protocol.
Further cultures not necessary unless patient identified by study staff to have VAE. (appendix A)
Study staff to be notified immediately of patient extubation from mechanical ventilation, discharge from intensive care unit or expiration.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Primary purpose
Allocation
Interventional model
Masking
120 participants in 2 patient groups