Status
Conditions
About
The goal of this observational study is collect data to evaluate the efficacy of the RMS system in monitoring, recording, and presenting respiratory function data to the user in participants scheduled for pulmonary function testing (PFT).
Participants will complete:
The TSS will continuously transmit sound data to an adjacent personal computer (PC) via Bluetooth Low-Energy (BLE). TSS trachea sound data will be recorded on the PC and then transmitted via a secure wireless network to an RTM cloud account that is HIPPA compliant.
Reference breathing data will be simultaneously recorded using an FDA approved hospital ventilator (Hamilton Medical, HAMILTON-C1) with a calibrated pneumotach, capnometer, and a tight-fitting face mask. This system accurately measures and records a spontaneously breathing patient's RR, TV, MV, and end-tidal carbon dioxide concentration.
Full description
Observational single-center data will be captured on up to 50 patients scheduled for pulmonary function testing (PFT) at Thomas Jefferson University Hospital's (TJUH) outpatient PFT facility. Plan is to record a complete data set from 30 patients. Breathing data for this study will be collected in the PFT facility by research personnel approximately 1 hour before or after the scheduled PFT test. Study personnel will adhere the TSS to the skin over the proximal trachea (midline above the sternal notch). The patient's neck will be examined prior to TSS attachment.
The TSS will continuously transmit sound data to an adjacent personal computer (PC) via Bluetooth Low-Energy (BLE). TSS trachea sound data will be recorded on the PC and then transmitted via a secure TJUH wireless network to an RTM cloud account that is HIPPA compliant.
Reference breathing data will be simultaneously recorded using an FDA approved hospital ventilator (Hamilton Medical, HAMILTON-C1) with a calibrated pneumotach, capnometer, and a tight-fitting face mask. This FDA approved system accurately measures and records a spontaneously breathing patient's RR, TV, MV, and end-tidal carbon dioxide concentration.
Study subjects will breathe through the HAMILTON ventilator's face mask, , capnometer, and pneumotach over a range of RR, TV, and MV with the data recorded on a computer. Each study subject will breathe with a normal RR, TV, and MV for approximately 5 minutes. Each subject will then breathe with a normal RR and a decreased tidal volume for approximately 60 to 80 seconds episode (shallow breathing), followed by a rest period (normal breathing). Each subject will then breathe with a normal RR and an increased tidal volume for approximately 60 to 80 seconds episode (deep breathing), followed by a rest period (normal breathing). This series of breathing will take approximately 7.5 minutes to record. Each subject will complete a second series of breathing using the identical methods. Study subjects will observe a computer screen that displays breath- by-breath tidal volume to help them maintain their tidal volume in the desired range during each recording.
Plan is to record and analyze approximately 280 breaths per study subject and 8,400 breaths from the 30 study subjects.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Loading...
Central trial contact
Breanna Atherton, BS; Jeffrey Joseph, DO