BACKGROUND:
Hepatocellular carcinoma is the fifth most frequent cancer in the world, with a diagnosis of more than 500,000 new cases per year. It is considered the third leading cause of cancer mortality and presents well-defined risk factors. Liver cirrhosis is the main risk factor for developing HCC, therefore screening programs in cirrhotic patients will allow the early diagnosis of this neoplasia. Despite this, most HCCs are diagnosed at a stage in which the application of curative therapies is no longer possible.
Hepatic transarterial chemoembolization (TACE) belongs to the arterially directed embolization therapies for the treatment of unresectable early-to-advanced hepatocellular carcinoma (HCC). It is the only therapy that has shown to improve survival in intermediate-stage HCC.
Drug-eluting beads (DEB)-TACE has shown to provide slow drug elution, reduced liver and systemic toxicity, increased local drug concentration, and tissue necrosis.
Aside from TACE, other transarterial options include bland embolization, or hepatic artery embolization (HAE), and transarterial radioembolization (TARE). All have an acceptable safety profile, and each has its associated procedural and peri-procedural complications. One potential complication that may occur during all embolization procedures is when the embolic material migrates outside of the desired treatment area, leading to non-target embolization (NTE). In fact, when collateral vessels are embolized, there is a risk that these may be feeders of non-target tissue or organs.
NTE following TACE in particular may lead to a double-layer problem: dangerous components affecting healthy tissue, one ischemic and one related to cytotoxicity from the chemotherapeutic agent, which may have clinical consequences, and potential incomplete treatment of the lesion (due to beads being "deviated" from target).
NTE is highly recognized, but often thought to be uncommon, and although different complications can be caused by it, there may appear to be no evidence of NTE during the intraprocedural imaging.
To avoid the complications due to NTE, apart from the importance of the pre-, intra- and post-procedural imaging, and the thorough study of the anatomical picture, the catheters/microcatheters should also be chosen with reason and care. In particular, selective catheterization should be achieved by placing the microcatheter tip as close as possible to the target, through the specific branch/branches supplying it.
However, even with the microcatheter selectively positioned in the vessel to be embolized, the risk of NTE might not be eliminated, since it could happen as a result of changes in flow dynamics that occur during embolization, particularly when the endpoint is stasis. These changes could result in reflux into non-target territories and, as such, might be better prevented with the use of microcatheters intended to reduce reflux. To this purpose, the use of a dedicated delivery device should be taken into consideration, in order to optimize and save time during the procedure.
Microcatheters are commonly used during most arterial embolization procedures, and as explained above, there is a strong rationale to use a reflux-control microcatheter - like Sequre - for DEB-TACE.
The main expectation is to achieve technical success with Sequre in all patients with a reachable target lesion, with the intent not only to minimize potential damage to surrounding tissue, but also to potentially deliver more treatment embolics, as all the beads are (re)directed towards the target.
The use of small diameter particles (100 micron-TANDEM ® spheres), induces superior tumor necrosis response (Urbano et al., European Journal of Radiology, 2020); with the synergistic effect of being administered through the SEQURE anti-reflux protection system, there is reason to believe that it will be possible to administer maximum doses of doxorubicin, while avoiding the occlusion of non-target arterial segments (SYNERGIC EFFECT).
STUDY PROPOSAL:
We propose a prospective observational study with data collection from a single center (Virgen de las Nieves University Hospital-Granada), for a period that ranges October 2020-December 2021. Here summarized the inclusion criteria and contraindications:
Inclusion criteria
Contraindications
Full description
Microcatheters are commonly used during most arterial embolization procedures. SEQURE® (Guerbet, France) is an innovative reflux-control microcatheter for peripheral embolization procedures which relies on flow dynamics as the foundation of its' mechanism of action. It can optimize the delivery of calibrated microspheres of specific sizes.
The device presents as a regular microcatheter with a non-tapered internal lumen for delivery of embolics (typically in suspension with contrast media), and 2 radiopaque markers (placed at a distance of 11 mm) at the tip (distal marker being at 0.5-1 mm from the point of exit).
Near the distal atraumatic tip, between the two markers, it has side slits (approximately 50 µm in width) which allow contrast media to exit radially in addition to the standard fluid flow through the distal tip of the catheter. The contrast media that comes out of the side-holes creates a turbulence in the space surrounding the slits, between microcatheter and vessel, which acts as a fluid barrier in that given portion. This fluid-dynamic barrier redirects the beads that tend to upstream reflux back to their flow-directed trajectory, preventing the beads from migrating in non-target areas, and thus eliminating, or strongly reducing, undesired non-target embolization (NTE).
Moreover, SEQURE® allows for haemodynamic arrangement to be preserved, by favoring and maintaining unaltered the natural blood flow. Under fluoroscopy, the radial flow of contrast media exiting through the side holes in the SEQURE® microcatheter can be seen, giving the physician indication of the vessel flow dynamics.
Lesions were identified and then treated supra-selectively with the microcatheter, with the embolic materials infused in a slow and controlled manner, in order to reduce reflux into non-target branches
HYPOTHESIS:
The use of 100 micron-TANDEM ® spheres will be studied with the synergistic effect of being administered through SEQURE® microcatheter anti-reflux protection system, which may allow the administration of the maximum doses of doxorubicin, avoiding the undesired oclussion of collateral arterial segments (SYNERGIC EFFECT) in a prospective study in patients with hepatocellular carcinoma early-intermediate stage.
PATIENT SELECTION:
Between October 2020 and December 2021, and expected number of 30 patients will be treated. The diagnosis of Hepatocellular carcinoma (HCC) will confirm in accordance with the American Association for the Study of Liver Diseases guidelines and HCC staged according to Barcelona Clinic Liver Cancer (BCLC) system.
DEB-TACE is indicated according to clinical practice guidelines referred above.
TECHNICAL PROCEDURE:
The embolization will be performed using TANDEM ® spheres (each cc of spheres can be loaded with up to 50 mg of doxorubicin. The maximum dose of doxorubicin used per procedure was 150 mg, corresponding to 3 cc of spheres) and SEQURE ® microcatheter in a super-selective tumour´s arterial position avoiding undesired arterial branches. The procedure will consider super-selective when the micro-catheter tip reached the tumour feeding artery. Each cc of microspheres will mix with 7-10 cc of non-ionic contrast medium. Embolization endpoint is considered vascular stasis or 150 mg of Doxorubin administered.
OBJECTIVES AND END-POINTS:
Efficacy:
Safety:
DATA COLLECTION AND FOLLOW-UP:
The information will be collected through the electronic database of the patient (includes laboratory data). Patients will be evaluated starting from the date of inclusion as candidates for TACE treatment, and followed up for 6 months (unless patients are lost to clinical follow-up or death occurs). Mortality data are obtainable by searching the electronic history of the patient.
Changes in the serum levels of aspartate aminotransferase (AST), alanine aminotransferase (ALT), Alkaline phosphatase (ALP), bilirubin, albumin and patients' complete hematological profile, as well as serum lipase and amylase levels within the 1st week post-TACE to detect pancreatitis, will be evaluated at the following time-points: pre-embolization, at one month, three months, and six months. Computed tomography (CT) or Magnetic Resonance Imaging (MRI) with contrast will be performed at the same time-points as just indicated; modified Response Evaluation Criteria in Solid Tumors ( m-RECIST) will be used, and images will be assessed by two experienced radiologists (any discordant interpretation will be solved through team consensus). All serologic toxicities would be classified according to common terminology criteria for adverse effects. On this scale, severe hepato-toxicity (grade 3) is classified as the increase of up to 5 times the normal limit of AST, ALT, FA or the increase of bilirubin level to over 3 g/dl.