SpO2 Measurement Depending on Different Pulse Oximeter Averaging Time Settings

C

Czech Technical University in Prague

Status

Completed

Conditions

Reduced Perfusion During SpO2 Measurements
Movement Artifacts During SpO2 Measurements

Treatments

Other: Artifact - ambient light
Other: Movement artifact - knocking
Other: SpO2 measurement
Other: Movement artifact - tremor
Other: Artifact - low perfusion

Study type

Interventional

Funder types

Other

Identifiers

NCT04933773
POAveragingTime21

Details and patient eligibility

About

The aim of this study is to characterize the noise in the SpO2 signal depending on different pulse oximeter averaging time settings.

Full description

Noise in the SpO2 signal is most often caused by artifacts, such as movement artifacts, reduced perfusion, additional light and others. These artifacts can be partially removed from the SpO2 signal by different pulse oximeter settings, specifically the sensitivity mode and averaging time. It is known that when a longer averaging time is set, the measured signal contains fewer artifacts, on the other hand there is an increased risk of missing short desaturation. The aim of this work is to characterize the noise in the SpO2 signal depending on the set averaging time of the pulse oximeter based on experimental data from healthy volunteers.

Enrollment

17 patients

Sex

All

Ages

18 to 30 years old

Volunteers

Accepts Healthy Volunteers

Inclusion criteria

  • healthy volunteers from the Czech Technical University

Exclusion criteria

  • pregnancy
  • severe cardiovascular conditions
  • injury to the upper limbs or hands that could affect the peripheral perfusion
  • diabetes
  • hypotension or hypertension

Trial design

Primary purpose

Other

Allocation

Randomized

Interventional model

Crossover Assignment

Masking

Single Blind

17 participants in 2 patient groups

Set averaging times: 2-4 seconds and 8 seconds
Experimental group
Description:
SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 2-4 seconds, the second pulse oximeter is set to an averaging time of 8 seconds.
Treatment:
Other: Movement artifact - knocking
Other: SpO2 measurement
Other: Movement artifact - tremor
Other: Artifact - low perfusion
Other: Artifact - ambient light
Set averaging times: 8 seconds and 16 seconds
Experimental group
Description:
SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 8 seconds, the second pulse oximeter is set to an averaging time of 16 seconds.
Treatment:
Other: Movement artifact - knocking
Other: SpO2 measurement
Other: Movement artifact - tremor
Other: Artifact - low perfusion
Other: Artifact - ambient light

Trial contacts and locations

1

Loading...

Data sourced from clinicaltrials.gov