Validation study to validate the SpO2 accuracy performance of the Sponsor's SpO2 sensors over the range of 70-100% SaO2 during non-motion conditions assessed by CO-Oximetry
Full description
The intent of this clinical study is to validate the SpO2 accuracy performance of the Sponsor's SpO2 sensors over the range of 70-100% SaO2 during non-motion conditions assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of the Sponsor's pulse oximeter with consideration to anticipated updates to ISO 80601-2-61:2025(E)
Enrollment
Sex
Ages
Volunteers
Inclusion and exclusion criteria
INCLUSION CRITERIA
Participants accepted in this study must meet all the following criteria:
EXCLUSION CRITERIA
Participant is considered as being morbidly obese (defined as BMI >39.5)
Compromised circulation (i.e., Raynaud's Syndrome), injury, open wound, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoos in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the sites utilized.)
Females who are pregnant
Females who are trying to get pregnant (with confirmation of positive urine pregnancy test unless the participant is known to be not of child-bearing potential)
Participants who have smoked in the last 2 days or participants who have refrained, for at least 48 hours, with COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow)
Participants with known respiratory conditions such as: (self-reported)
Participants with self-reported heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)
Self-reported health conditions as identified in the Health Assessment Form
Participants with known clotting disorders (self-reported)
Participants with severe contact allergies to standard adhesives, latex, silicone or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self-reported)
Participants with severe allergies to iodine (only applicable if iodine is used)
Participants with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocaine),
Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio <0.4)
Unwillingness or inability to remove nail polish or artificial nails from test digits.
Participant received intravascular dye within the past 48 hours (e.g. Indocyanine green, methylene blue, dyes used in cardiac output monitoring)
Surgical hardware in pathway of Device Under Test
Other known health conditions should be considered upon disclosure in health assessment form.
Participants with uneven skin tone at the sensor site or at the forehead
24 participants in 4 patient groups