Status
Conditions
About
The product is comprised of software that aids in automated guided ultrasound image acquisition for novice users (medical professional without experience in performing echocardiography, e.g. nurses, physicians assistants, medical doctors, medical student etc.).
The software provides real-time feedback during image acquisition to optimize image capture and determine whether image quality is sufficient for medical evaluation. The software is designed to be used with any echocardiographic ultrasound system. The Philips Lumify hand-held ultrasound device will be used for this study.
Full description
Study Design:
This pilot study is a single center, multi-reader multi-case study, primarily designed to obtain the quality of the echocardiography exam, obtained by novice users when using the USEG software.
Eligible subject will undergo the examination twice on a single study visit : once by one of novice users when using the USEG software and once by one of the expert sonographers without the aid of the software.
Each novice user will perform the examination on approximately 30 subjects (depending on the number of novice user, with the aim that the number of examinations by each novice will be the same).
The assignment of the novice user and expert sonographer to subjects and their order within a subject will be based upon their availability once the subject's is enrolled. In case of more than 1 available novice, the exam will be assigned to the novice with the least exams performed so far.
At least 10 subjects will be enrolled from each of the following three BMI groups= < 25; =25 to <30; => 30.
Study Procedures:
After an informed consent is obtained, screening was completed and eligibility has been confirmed, each subject will undergo the examination twice on the day of visit (Day 1): once by a novice user and once by an expert sonographer.
In case of any administrative delay, the investigator may address the issue with the sponsor, and upon approval, the second examination may be postponed in up to 96 hours.
The novice user will obtain at least 10 clips using the UltraSight guidance software for 10 pre-defined views from each patient, and an expert sonographer will obtain 10 clips for the same at least 10 views without the aid of the UltraSight software.
The order in which the novice user and expert sonographer performs the examination per patient will be based on their availability once the subject's is enrolled. In case of more than 1 available novice, the exam will be assigned to the novice with the least exams performed so far.
Adverse events will be monitored throughout the visit.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Emergency (non-elective) admission within 24 h prior to participating in the study
Female subjects who are pregnant (WOCBP* will perform a urine pregnancy test)
Unable to lie as required in all the classic positions for standard TTE exam: supine on back / left decubitus.
Subjects who currently participate a clinical trial, involving interventional cardiac devices
Subjects who have prior Echo exam with description of Low/Poor quality exam in the echo report
Subjects with BMI above 40
Subjects experiencing a known or suspected acute cardiac event
Subjects with severe chest wall deformity as per previous medical records and physical examination
Subjects who have undergone pneumonectomy
Subjects whose anatomy does not lend itself to yield diagnosable echocardiography clips (i.e. situs inversus with dextrocardia, single ventricle anatomy due to congenital heart defect, etc.)