PURPOSE OF EVALUATION: This study will verify & validate the acceptability of Egret's extended use product as compared to the existing daily change-out product. Based on an extensive in-vitro study, Egret's product confirmed significant kill for both positive and negative gram bacteria. In addition to meeting all of the current product requirements, total aerobic bio-burden on the catheters will be tested at various intervals for up to 10 days of use. The current used product will be used as a control group that is to be changed out daily as per current policy. The control group will be tested for total aerobic bacteria counts on the last day of use.
Full description
Protocol
Title: Extended Use In-line Endotracheal Suction Catheter w/BIOSAFE
Protocol Number: RT 2008-001
Study Director: Richard C. Dowdy
Study Director's Address and Method of Contact:
Egret Medical Products Inc. Richard C. Dowdy 2713 Industrial Lane Garland, TX 75041 (214) 291-0238 Phone (817) 991-7459 Cell (214) 291-0243 Fax rdowdy@egretmedical.com
I. PURPOSE OF EVALUATION: This test will verify & validate the acceptability of Egret's extended use product as compared to the existing Ballard's daily change-out product. In addition to meeting all of the current product requirements, total aerobic bio-burden on the catheters will be tested at various intervals for up to 10 days of use.
II. BACKGROUND: Prior to this evaluation, a baseline study was conducted at Medical City to determine the existing total aerobic bio-burden after one day of use, using 25 randomly selected Ballard catheters as per the attached protocol. The data from the baseline study were used to establish an appropriate sample size for the Extended Use Protocol. This data gathered under this Extended Use Protocol will be compared to the baseline data to determine the effectiveness of Egret's antimicrobial properties during extended use.
III. POTENTIAL RISK TO PATIENT: Closed endotracheal suction catheters are intended to maintain a hermetical seal. However, such catheters have a propensity to build up bio- film on the tip area and require saline rinsing and cleaning after use. Egret's catheter is compounded with a non-leaching antimicrobial to address this issue. There is the possibility that excessive patient secretions can over saturate any closed suction catheter. Egret has addressed this concern by incorporating a unique squeegee-and-filter design that facilitates tip cleaning and prevents debris from migrating from the cleaning chamber to the sleeve area.
IV. BENEFTS OF THE EGRET MEDICAL DEVICE
V. PRODUCT to BE EVALUATED:
Product Code Description Size 454 Egret 14 Fr. T-Piece - Extended use 14 FR/4.6 mm 54 cm (21.75") length
VI. INVESTIGATOR David Barton, BA, RRT, RCP
VII. SPONSOR Egret Medical Products Inc.
VIII. AFFLICATION OF APPLICANT Egret Medical Products Inc.
IX. FUNDING AGIENCIES Egret Medical Products Inc.
X. ANTICIPATED TEST DATES April 2008-May 2008
XI. METHODOLOGY:
XII. DATA TO BE COLLECTED: After product has been used, the healthcare participants will complete a written questionnaire concerning the test device as per attached.
XIII. DELIVERY OF PRODUCT: Available test product will be delivered to use-test site as per Medical City instructions.
XIV. RECOVERY OF PRODUCT: The investigator will be responsible for gathering all unused test product and returning it to Egret.
XV. SPECIAL INSTRUCTIONS: Participants will be briefed on the specific features of Egret's extended use product to be tested at their facility.
XVI. TEST FACILITATORS: Egret will provide on-site qualified personnel to assist with resources as required by Medical City.
XVII. APPROVALS
Study Director: Richard C. Dowdy Date: Title: Vice President
The Study Director is responsible for ensuring that no data for use in the Final Report (Outputs) has been generated before protocol approval.
Approvers: Egret Medical Products Inc. Management
Signature(s) indicate that this protocol has been reviewed and understood in its entirety and meets all criteria specified in Section 4.1, Protocol Design and Approval Policy.
Approver: Date: Title:
Approver: Date: Title:
Approver: Date: Title:
Upon approval of the protocol, the original is to be filed at the administrative offices of Egret Medical Products Inc. The study director should retain a copy of the protocol until the report is completed
Endotracheal Suctioning Catheter Use Test Questionnaire
Clinician's Name: Date:
Where Used: Oral Intubation 0 or Trach 0
Product usage questions Not Same as Improved Acceptable Current Use of Change-out label
0 0 0
Connection to suction tubing
0 0 0
Suction Valve - Ease of use
0 0 0
Gauge and feel of sheath material
0 0 0
Valve Locking Safety Feature
0 0 0
Ease of connection to trach or ET tube
0 0 0
Catheter positioning and ease of manipulation
0 0 0
Communication\visibility of depth markings
0 0 0
Use of saline port
0 0 0
Tip cleaning process and result
0 0 0
Product flexibility to prevent unnecessary ET movement?
0 0 0
Did sheath remain clean and dry? 0 0 0
Affect of heat on product function? 0 0 0
Perceived weight of product 0 0 0
Comfort to patient
0 0 0
Did end cap remain secure?
0 0 0
Ease of disconnect after use?
0 0 0
Did this product meet your expectation? 0 0 0
Additional Comments:
Thank you for your help
Enrollment