Study to Verify Proper Detection of Supraventricular Tachyarrhythmia With Single-Lead Dual-Chamber Implantable Cardioverter-Defibrillators (ADRIA)

B

Biotronik

Status and phase

Completed
Phase 4

Conditions

Tachyarrhythmia

Treatments

Device: Implantable cardioverter-defibrillator

Study type

Interventional

Funder types

Industry

Identifiers

Details and patient eligibility

About

A unique single-lead dual-chamber implantable cardioverter-defibrillator (ICD) system (produced by Biotronik, Germany) features a conventional electrode in the ventricle (anchored in the ventricular apex) and a floating electrode (ring on the lead body) in the atrium, capable of sensing atrial electrical signals. The purpose of this study is to determine whether this system detects a supraventricular tachyarrhythmia (e.g. atrial fibrillation, atrial tachycardia) in the equivalent manner as conventional dual-lead dual-chamber ICDs using two separate electrodes anchored in the ventricle and in the atrium, respectively.

Full description

Supraventricular tachyarrhythmia (SVT) is the main cause of inappropriate therapy in patients with single-chamber implantable cardioverter-defibrillators (ICDs). To minimize inappropriate shock delivery, ICDs should sense both atrial and ventricular intracardiac signals. Dual-chamber ICDs are used for this purpose, but are associated with increased postoperative complications due to the implantation of a separate atrial lead. It has to be shown that a novel single-lead dual-chamber ICD-system with enhanced SVT discrimination can achieve the same specificity in discriminating SVT episodes as conventional dual-chamber ICD, without the disadvantage of the implantation of several leads. In this study patients eligible for dual-chamber ICD therapy who do not need atrial pacing will receive either a single-lead dual-chamber ICD (Belos A+ and Kainox A+ electrode) or a dual-lead dual-chamber ICD (Belos DR). SMART detection algorithm will be used in both study groups for discrimination between atrial and ventricular tachyarrhythmias. Atrial tachyarrhythmia episodes are facultatively induced in both groups at implantation or predischarge via a bipolar stimulation catheter. Induced episodes and the corresponding ICD intervention (detection and therapy or inhibition of therapy) are documented. Follow-ups are scheduled for 1, 3, 6 and 12 months after implantation. ICD intervention, particularly related to spontaneous SVT episodes, will be evaluated based on ICD diagnostic memory data interrogated at follow-up controls.

Enrollment

260 estimated patients

Sex

All

Ages

18+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • ICD indication according to American College of Cardiology/American Heart Association (ACC/AHA) guidelines.
  • Informed consent

Exclusion criteria

  • Permanent atrial fibrillation
  • Requirement for atrial pacing
  • Patient is underage
  • No signed patient agreement
  • Patient life expectancy under 6 months

Trial design

Primary purpose

Treatment

Allocation

Randomized

Interventional model

Parallel Assignment

Masking

None (Open label)