Status
Conditions
Treatments
About
This study compares two intubation boxes, that is the Taiwan "Aerosol Box" versus the UMMC "Intubation Box". The importance of this intubation box has come to light recently in view of the COVID-19 pandemic and the fact that intubation is an aerosol-generating procedure. Risks of the healthcare worker attending to the airway of COVID-19 patients is high and the intubation box aims to minimize that and reduce contamination of the environment. The original design of the "Aerosol Box" has limitations and is not easily used by intubators. This study evaluates the clinical usefulness of an innovation of the "Aerosol Box" design that is made to be more ergonomic and increases the rate of successful intubation. Anaesthetists with more than 5 years of clinical experience in intubating airways as well as more than 20 successful intubations using videolaryngoscopy , will be available to participate. This study will employ manikin and will be a randomized cross-over trial, conducted in UM.
Full description
Background:
Intubation is classified as an aerosol-generating procedure (AGP). Intubation in a confirmed COVID-19 patient with a high viral load poses a high risk of exposure to health care workers (HCW). To reduce this risk, the HCW and their assistants are advised to don full personnel protective equipment (PPE) with a powered air purifying respiratory (PAPR) during intubation. In view of this concern, this procedure is highly recommended to be done in a negative pressure room to control the spread of aerosolizing particles in the room. A first-pass success in intubation is also crucial to minimize the risk of infection to health care workers involved.
Introduction :
An Aerosol Box (from here on known as Box A) was recently designed by Dr. Lai Hsien Yung from Taiwan which adds extra protection to the intubator and the surrounding environment. The Aerosol Box is a transparent box made of acrylic or transparent polycarbonate sheet, designed with an opening on one side allowing it to fit over the patient's chest and neck, while the opposing side has two holes through which the intubator can insert their hands through. It was shown that this barrier enclosure during intubation protects the laryngoscopist. A recent study simulated a patient with a cough during intubation with and without the Aerosol Box. With the box, it was demonstrated that the simulated cough resulted in contamination of only the inner surface of the box, the laryngoscopist's gloves and gowned forearms. Examination of the laryngoscopist and the room with ultraviolet light after the simulated cough also showed no macroscopic contamination outside the box. In contrast, intubation without the box demonstrated that the laryngoscopist's gown, gloves, face mask, eye shield, hair, neck, ears, and shoes were all contaminated.
Feedbacks from our colleagues stated that intubation with Box A is slightly difficult and may cause delay in intubation. Canelli et al also concluded this observation. The investigators, therefore, innovated the design of Box A to facilitate the laryngoscopist, known as the UMMC Intubation Box (from here on known as Box B).
Enrollment
Sex
Volunteers
Inclusion criteria
Exclusion criteria
Primary purpose
Allocation
Interventional model
Masking
30 participants in 2 patient groups
Loading...
Central trial contact
MOHD F ZAINAL ABIDIN, MAnaes; SHAIRIL R RUSLAN, MAnaes(UM)
Data sourced from clinicaltrials.gov
Clinical trials
Research sites
Resources
Legal