Status
Conditions
Treatments
About
This is a prospective, single arm, early feasibility study designed to evaluate the safety and device performance of The RenalGuard System in the management of patients admitted with signs and symptoms of congestive heart failure who require diuresis for the treatment of volume overload.
Full description
This is a prospective, single arm, early feasibility study designed to evaluate the safety and device performance of The RenalGuard System in the management of patients admitted with signs and symptoms of congestive heart failure who require diuresis for the treatment of volume overload. Subjects will be referred by their physician if he/she feels that the subject could benefit from observed diuresis in a hospital setting.
The device being utilized is called The RenalGuard® System (The System). By default, The System infuses a volume of hydration fluid equal to the volume of urine output. This is known as matched hydration. In addition, the clinician has the ability to adjust the matched setting such that RenalGuard can infuse hydration fluid to achieve a positive fluid balance (e.g. +100 ml/hr.) or a negative fluid balance (e.g. -100 ml/hr.). The purpose of this balanced fluid replacement is to prevent hypovolemia that may lead to hypotension and vital organ dysfunction, or fluid overload that may lead to shortness of breath in subjects in whom high urine output is desired.
All patients will be treated with The RenalGuard System for a minimum of 24-hours, and possibly longer, at the investigators discretion. They will be followed for 30 days post discharge. The protocol will enroll up to 40 subjects to develop the clinical algorithm required to optimally and safely perform removal of volume. The study will be conducted in up to 5 sites in Europe.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Patient ≥18 years old (with body weight ≤160 kg) who is able to provide informed consent
Subject is diagnosed and hospitalized with acute decompensated heart failure. ADHF is defined as including all of the following measured at any time between presentation and the end of screening:
Subject with pre-existing chronic renal failure (impaired renal function) defined as an eGFR between presentation and enrollment of ≥ 25 and <90 mL/min/1.73m2, calculated using the MDRD equation.
Subject has agreed to all follow-up testing.
Exclusion criteria
Primary purpose
Allocation
Interventional model
Masking
40 participants in 1 patient group
Loading...
Central trial contact
Andrew Halpert
Data sourced from clinicaltrials.gov