Temperature Monitoring With InnerSense Esophageal Temperature Sensor/Feeding Tube After Birth Through Stabilization in VLBW Infants

Duke University logo

Duke University

Status

Completed

Conditions

Body Temperature

Treatments

Device: Philips InnerSense Esophageal Temperature Sensor/Feeding Tube
Device: Standard of care feeding tube

Study type

Interventional

Funder types

Other
Industry

Identifiers

NCT02311972
Pro00051454

Details and patient eligibility

About

Very low birth weight (VLBW) infants (<1500 grams at birth) and other low birth weight infants experience hypothermia after birth and through stabilization (first 24 hours of life) due to an inability to keep warm through metabolic heat production while experiencing heat loss during care.

The investigator hypothesizes that inserting an InnerSense oral gastric tube (Philips Healthcare) with its imbedded thermistor and attaching the tube to a temperature monitor with a digital temperature display will enable care-providers to monitor continuous body temperature and prompt them to provide warmth and prevent heat loss, thus preventing hypothermia in these infants for the first 24 hours of life. The investigator predicts infants in the intervention group will have warmer axillary temperatures upon admission to the NICU and at 1, 4, and 8 hours of age. Mothers in non-active labor will be sought for consent to enroll their VLBW infant into this study after the infant's birth. 160 VLBW infants and other low birth weight infants 1000-2000 grams will be randomized to the intervention group (placement of InnerSense oral gastric tube with thermistor to a Squirrel monitor for continuous digital temperature display, immediately after birth through 24 hours of life) or the control group (standard delivery room care and stabilization care). The study staff plans to enroll 180 infants (90 per group) to allow for screen failures. Axillary temperatures of all infants will be recorded at NICU admission, 1, 4, 8 and 24 hours of age. Further, the study staff intends to enroll approximately 300 Intensive Care Nursery providers who care for an InnerSense study infant and wish to complete a provider satisfaction survey. These providers will be asked to provide consent prior to completing the survey and their names will be recorded on a separate study enrollment log.

Infant demographic data will be compared to make sure groups are not significantly different. Temperatures from the intervention group will be compared to the control group using a student t test at each time point. Satisfaction questionnaires will be summarized and described. The InnerSense oral gastric tube with temperature monitoring is FDA approved for use and is being marketed commercially by Philips Healthcare. VLBW infants have oral gastric tubes placed as standard of care for feeding, and intermittently in the delivery room when positive pressure ventilation is necessary for respiratory stabilization. There are no additional safety risks to the intervention of this study; however, a safety/data monitoring committee will be formed and meet quarterly to monitor the trial.

Enrollment

52 patients

Sex

All

Ages

1 minute to 24 hours old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • Parents of eligible infants must be enrolled prior to, or shortly after delivery as the InnerSense esophageal temperature sensor/feeding tube will be inserted post delivery resuscitation

Exclusion criteria

  • Infants are excluded if parents were not consented prior to, or shortly after delivery
  • Infants are not eligible if they have major congenital anomalies including gastrointestinal or facial abnormalities
  • Infants with major cardiac anomalies are also excluded

Trial design

Primary purpose

Prevention

Allocation

Randomized

Interventional model

Parallel Assignment

Masking

None (Open label)

52 participants in 2 patient groups

Control - Standard of Care
Active Comparator group
Description:
Infants in the standard of care group will receive no study interventions other than study recorded axillary temperatures on admission, and at 1, 4, 8 and 24 hours. Infants will receive standard delivery room stabilization and NICU stabilization. Infants in the control group will receive a feeding tube as standard of care, and the standard issue feeding tube used in the Intensive Care Nursery does not have the capability to record or display esophageal temperatures. Nurses will record an axillary temperature upon admission to the NICU, and at 1, 4, 8 and 24 hours of age for each infant in both groups on a data sheet at the bedside. These data will be entered into a RedCap data base created for the study.
Treatment:
Device: Standard of care feeding tube
Philips InnerSense Esophageal Temperature Sensor/Feeding Tube
Active Comparator group
Description:
The nurse will insert an InnerSense temperature sensor/feeding tube per normal standards as soon after birth as possible, during delivery room stabilization. Nurses will record an axillary temperature upon admission to the NICU, and at 1, 4, 8 and 24 hours of age. The InnerSense tube will be attached to the infant's bedside monitor to continuously display esophageal temperatures. The tube will stay in place until the infant is 24 hours of age. Central body temperature will be displayed continuously for care-providers in the delivery room, through transport from the birthing center to the NICU and through stabilization. Infants will have a thermistor placed on the abdominal skin with standard skin tape once admitted in the NICU. This thermistor will be attached to a Squirrel SQ2010 (Grant Instruments) temperature monitor/data logger to collect abdominal temperatures every minute for the first 24 hours of life. All temperatures entered into the study database.
Treatment:
Device: Philips InnerSense Esophageal Temperature Sensor/Feeding Tube