Aortic Stenosis

Treatments

Device: Transcaval closure device (TCD)

Study type

Interventional

Funder types

NIH

Identifiers

NCT03432494
18-H-N045
999918045

Details and patient eligibility

About

Background:

Transcatheter aortic valve replacement (TAVR) is a non-surgical alternative to standard surgical aortic valve replacement. Not all patients are eligible for TAVR using standard access through the artery in the groin, because the artery is too small or too diseased. In this study, TAVR is performed using a new technique called transcaval access. The catheter is placed in the artery deep in the body by crossing through the wall of a deep vein. The connection between that vein and the aorta is closed with a new metallic device they are testing. This is called a transcaval closure device (TCD).

Objective:

To test the safety and early feasibility of closure of transcaval aortic access sites using the TCD after TAVR.

Eligibility:

Adults ages 21 and older undergoing TAVR for whom the procedure cannot be performed safely by the standard artery approach

Design:

Participants will be assessed by heart experts including cardiologists and surgeons.

Participants will have TAVR by the transcaval approach. A small catheter will be passed between the largest vein in the body and the nearby largest artery (aorta), inside the abdomen. Through this catheter, the TAVR will be implanted in the usual way.

After, doctors will implant the TCD by catheter to close the hole made in the aorta.

Participants will be X-rayed. A dye will be injected to view the TCD device.

Participants will get standard TAVR care afterwards. They will have physical exams, blood tests, and scans.

Participants will have a follow-up scan within 1 month and after 12 months.

Participants will have follow-up visits and phone calls 6 and 12 months after the procedure.

Full description

Transcaval access to the abdominal aorta from the neighboring inferior vena cava (IVC) enables transcatheter aortic valve replacement (TAVR) in patients not eligible for femoral artery access. Currently the procedure is performed using devices, off-label, designed and marketed to close holes inside the heart and great vessels, manufactured by Abbott St Jude (Amplatzer Duct Occluder and Amplatzer Muscular VSD Occluder). Because these Amplatzer occluders are not designed to close transcaval access sites, they may not completely prevent bleeding.

This is an early feasibility study (EFS) evaluation of a purpose-built closure device for transcaval access. The device, the Transcaval closure device (TCD) will be evaluated for safety and performance to close transcaval access sites in patients ineligible for femoral artery access for TAVR.

Enrollment

12 patients

Sex

All

Ages

21+ years old

Volunteers

No Healthy Volunteers

Inclusion and exclusion criteria

  • INCLUSION CRITERIA:
  • Consents to participate in this study and all related clinical follow-up procedures
  • Adults age greater than or equal to 21 years
  • Undergoing transcatheter aortic valve replacement (TAVR) based on the clinical assessment of the multidisciplinary heart team
  • Ineligible for femoral artery access for the selected transcatheter heart valve (THV) according to the THV manufacturer instructions for use. (Note Corevalve Evolut R: greater than or equal to 5mm; Evolut PRO 23, 26, 29 mm valves and Evolut R 34 mm: greater than or equal to 5.5 mm; Edwards Sapien 3, 23-26mm: 5.5 mm; Edwards Sapien 3, 29mm: 6.0mm; Anatomic ineligibility also considers patient-specific pattern of iliofemoral calcium and tortuosity.)
  • Eligible for transcaval access based on Core lab analysis of the baseline abdomen/pelvis CT indicating a calcium-free target window on the abdominal aorta; a target greater than or equal to 15mm from the lowest main renal artery or aorto-iliac bifurcation; no important interposed structures; a projected intravascular centerline distance from the lower femoral head to the target at least 5cm less than the intended THV introducer sheath; patient celiac or superior mesenteric artery
  • Aorta diameter greater than or equal to 11mm at the target crossing site
  • Concordance of the study eligibility committee

EXCLUSION CRITERIA:

  • High risk features on baseline CT including porcelain aorta (confluent calcification); pedunculated aortic atheroma; or leftward aortic angel greater than or equal to 20 degrees with regard to vertical.
  • Renal dysfunction limiting follow-up contrast-enhanced CR (estimated glomerular filtration rate, eGFR less than 30mL/min/1.73m^2 if not already on renal replacement therapy)
  • Pregnancy or intent to become pregnant prior to completion of all protocol follow-up procedures

Trial design

Primary purpose

Treatment

Allocation

N/A

Interventional model

Single Group Assignment

Masking

None (Open label)

12 participants in 1 patient group

Transcaval access and closure with the transcaval closure device (TCD) test article.
Experimental group
Description:
All participants undergo transcaval access for transcatheter aortic valve replacement (TAVR) followed by implantation of the transcaval closure device (TCD).
Treatment:
Device: Transcaval closure device (TCD)

Trial documents