ClinicalTrials.Veeva

Menu

Performance of the L-PLUS BGMS by Persons With Diabetes Mellitus

Ascensia Diabetes Care logo

Ascensia Diabetes Care

Status

Completed

Conditions

Diabetes Mellitus

Treatments

Device: L-PLUS meter

Study type

Interventional

Funder types

Industry

Identifiers

NCT04033315
GCA-PRO-2019-002-01

Details and patient eligibility

About

L-PLUS (Lightning PLUS) is the code name of a blood glucose monitoring system (BGMS) consisting of a new meter and marketed CONTOUR PLUS® test strip. This clinical trial will assess the performance (accuracy) of the L-PLUS meter by lay users with diabetes enrolled as subjects in the study, and by health care professionals (also called study staff).

The trial will follow the requirements and procedures described in ISO 15197:2013 International Standard.

Full description

This study is designed to satisfy the ISO 15197:2013, Section 8 requirements.

The study will be conducted at a minimum of one clinical site and enroll a total of approximately 130 persons with diabetes.

For this study, there must be at least 120 evaluable results each for fingerstick and AST self-test results, as well as for staff testing subject fingertip and venous blood using the L-PLUS BGMS. Subject enrollment will continue until the required evaluable samples are collected.

The subjects will be 18 years and older. Each subject will make one visit, lasting approximately 1 hour at the clinical site. All subjects will complete the informed consent process before performing any study procedures.

All Subjects will be assigned a L-PLUS meter and one Contour Plus test strip lots. The untrained subjects will learn to use the L-PLUS meter by reading the instructional materials. Each subject will perform one fingerstick self-test, one palm test, and staff will perform a fingerstick test on the subject.

Subjects will be given a venipuncture and the venous blood will be tested by the study staff with the L-PLUS system. All subject capillary and AST self-test results and study staff fingerstick results will be compared to YSI fingerstick results. Venous blood meter tests will be compared to YSI results of the corresponding venous blood. Hematocrit will be measured for all subjects. All Subjects will then complete a questionnaire (Questionnaire 1) to provide feedback on the meter system and instructions for use.

Enrollment

130 patients

Sex

All

Ages

18+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • Males and females, 18 years of age and older
  • People with diabetes (PWD) - type 1 or type 2
  • Ability to speak, read and understand English. Subjects must demonstrate ability to read a paragraph from the first page of the UG to qualify for the study.
  • Willing to complete all study procedures

Exclusion criteria

  • Hemophilia or any other bleeding disorder
  • Pregnancy (self-reported)
  • Physical, visual or neurological impairments that would make the person unable to perform testing with the BGMS.
  • Previous participation in a blood glucose monitoring study using the Ascensia's Contour Next (Lightning Next) and Lightning PLUS (L-PLUS or Contour Plus ELITE) BGMS.
  • Working for a medical laboratory, hospital or other clinical setting that involves training on or clinical use of blood glucose monitors.
  • Being in this trial during or less than two days since xylose absorption testing (Xylose in the blood is known to cause interference).
  • Working for a competitive medical device company, or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company.
  • A condition, which in the opinion of the investigator or designee, would put the person or study conduct at risk. The reason for exclusion will be clearly documented by investigator or designee on the subject disposition form.

Trial design

Primary purpose

Supportive Care

Allocation

N/A

Interventional model

Single Group Assignment

Masking

None (Open label)

130 participants in 1 patient group

Untrained Subjects with Diabetes use the L-PLUS BGMS
Experimental group
Description:
Untrained subjects with Diabetes test capillary fingerstick and palm blood (and study staff test subject fingerstick blood) using the L-PLUS BGMS. All blood glucose results are compared to reference method results obtained with subject capillary plasma. Also, study staff test venous blood and blood glucose results are compared to reference method results obtained from subject venous plasma.
Treatment:
Device: L-PLUS meter

Trial documents
1

Trial contacts and locations

1

Loading...

Data sourced from clinicaltrials.gov

Clinical trials

Find clinical trialsTrials by location
© Copyright 2026 Veeva Systems