The ExVent is an optional accessory to the O2Vent Optima MAD and provides oral Expiratory Positive Airway Pressure (EPAP). Oral EPAP with the ExVent is designed to provide upper airway support via similar mechanisms of action of nasal EPAP devices in commercial distribution, e.g., passive dilatation of the airway, which reduces flow limitation. Nasal EPAP devices are in commercial distribution as stand-alone therapies for the treatment of OSA. The oral EPAP provided by the ExVent accessory is designed to augment the OSA therapy provided by the O2Vent Optima.
Full description
Title Use of the ExVent Accessory with the O2Vent Optima Oral Appliance for the Treatment of Obstructive Sleep Apnea
Investigational Device Oventus Medical ExVent Accessory (inserted into the O2Vent Optima device)
Investigator Study Sites Up to 3 investigational sites
Regulatory Status This is a non-significant risk investigational device exemption (IDE) clinical trial. FDA approval is not required, but IRB review and approval are required.
Study Design Prospective, open label, single-arm multicenter study Sample Size Up to thirty (30) subjects will be enrolled to yield a minimum of 20 evaluable subjects (mITT subjects).
Study Objective To demonstrate that the ExVent Accessory (inserted into the O2Vent Optima device) is safe and effective in treating mild to moderate OSA Patient Population Patients who have been diagnosed with mild to moderate OSA (i.e., AHI ≥ 5 and ≤ 30)
Study Overview Study Design/ Duration:
This is a prospective, open-label, single-arm study conducted at up to 3 sites. Site participation is expected to be approximately 6 months from first subject enrollment until completion of the last study visit.
Screening Phase
Home Use Phase
PSG Night Subjects who successfully complete* the Home Use Phase will have an in-lab PSG sleep night while using the O2Vent Optima + ExVent
Primary Effectiveness Outcome Measure Change in AHI between baseline AHI and the AHI measured during the in-lab PSG sleep night using the O2Vent Optima + ExVent.
Primary Effectiveness Endpoint Statistically significant change in AHI with use of the O2Vent Optima + ExVent compared to baseline AHI value (i.e., no treatment).
Safety Outcome Measure Adverse events will be summarized descriptively in terms of type and frequency
Statistical Analysis This clinical trial is designed to determine whether the concomitant use of the O2Vent + ExVent devices results in a statistically significant reduction in AHI from the baseline value for each subject. A paired t-test will be used for this comparison.
A sample size calculation is based on data from a comparable group of patients (Lai, 2019), where it was determined that the mean change in AHI from baseline was -8.37, with a standard deviation of the paired differences of 7.73.
Setting alpha=0.05 and power=0.80, the calculated sample size is 9 subjects. Because of the imperfect predictability of historical data, to ensure adequate power this study will enroll 20 evaluable patients.
The mITT population will be used for this analysis**
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Inclusion Criteria:
Exclusion Criteria:
Primary purpose
Allocation
Interventional model
Masking
20 participants in 1 patient group