aim of our study to validate electrical cardiometry readings against same readings by transoesophageal echo. we will validate stroke volume (SV) readings stroke volume variation(SVV), index of contractility.
Full description
Induction of anaesthesia will be by using propofol (2mg.kg) IV, fentanyl (1-2μg. kg) IV and atracurium (0.5 mg.kg) IV. Anaesthesia will be maintained with Sevoflurane adjusted between 1-2% in an air ⁄ oxygen mix (FiO2 0.6), fentanyl infusion at 1-2 μg.kg/h and atracurium infusion at 0.5 mg.kg/ h. Mechanical ventilation will be provided by using a Dräger anaesthesia machine (Dräger Primus®, Germany) using a tidal volume of 6-8 ml.kg with the respiratory rate adjusted to maintain the PaCO2 between 4-4.6 kilopascal (kPa) and positive end expiratory pressure (PEEP) of 5 cmH2O. All patients will be monitored for five lead ECG, peripheral oxygen saturation, noninvasive and invasive arterial blood pressure, temperature, end-tidal carbon dioxide tension, hourly urinary output, and central venous pressure (CVP). A 7-Fr triple lumen CVP catheter (Arrow International Inc, Reading, PA, USA) will be inserted into the right internal jugular vein.
All patients will receive 6ml /kg/h Ringer acetate solution as a maintenance intraoperative fluid. If SVV is more than 15%, the patient will be considered as fluid responder and will receive a 250-ml bolus of albumin 5% to maintain SVV ≤15%. Blood transfusion will be given based on a hemoglobin level (< 7 g/dl). Norepinephrine will be administered if the mean arterial pressure was less than 70 mmHg. Epinephrine will be administered if mean arterial blood pressure was less than 70 mm Hg and the cardiac index was less than 2.5 L/min/m2 despite sufficient volume infusion, to maintain a target cardiac index of 2.5-3.0 L/min/m2
Electrical cardiometry (EC) Examination:
ECG electrodes will be placed on the bare skin of patients in the following positions:
Patient characteristics of sex, age, height and weight will be entered into the monitor.
Clocks on the EC and trans esophageal echo (TEE) machine will be synchronized prior to data collection.
The EC monitor will be programmed to capture moving averages of variables based on the previous 10 cardiac cycles and to record those averages every10 s.
Only EC data with a signal quality index (SQI) of >70 will be included in the analysis.
Echo Examination:
A single LVOT diameter will be measured for each patient as the distance between the inflection points at the base of the aortic valve cusps from the left ventricle long axis view during systole.
Assuming a circular cross section, the LVOT area will be calculated from the LVOT diameter as:
π X (LVOT diameter/ 2)2 = (LVOT diameter)2 X 0.785
Pulse wave Doppler sampling cursor will be placed in the middle of the LVOT immediately proximal to the aortic valve in 5 chamber deep transgastric view.
The sonographer manually will trace the velocity-time envelope.
SV values will be identified for 10 cycles and averaged to obtain SV-Doppler max and SV-Doppler min. The mean SV (SV-Doppler mean) will be calculated as (SV-Doppler max - SV-Doppler min)/2.
SVV-Doppler will be calculated as (SV-Doppler max - SV-Doppler min)/SV-Doppler mean.
Peak aortic velocity, time to peak and mean acceleration will be calculated.
Mini Fluid challenge:
Mini fluid challenge will be done by 150 cc crystalloid over 1 min
2 sets of measurement of SV and SVV will be obtained