Validation of ECG Waveform and Arrhythmia Diagnosis Concordance for Pediatric Patients Using ECG Monitoring Patch

M

Mezoo

Status

Enrolling

Conditions

Arrhythmias, Cardiac

Treatments

Device: Electrocardiogram measurements by holter device and patch-type electrocardiographic at the same time

Study type

Interventional

Funder types

Other

Identifiers

NCT05756309
CDV-2022-02

Details and patient eligibility

About

This study will verify the 1-channel patch-type ECG's arrhythmia diagnostic concordance rate and electrocardiogram measurement performance in pediatric patients by collecting ECG signals from patch-type ECG and Holter device simultaneously for data analysis.

Full description

[How to conduct research] This study targeted more than 29 days old and less than 19 years old. Written consent is obtained from those who need a Holter examination among outpatients or inpatients at the Pediatrics of Samsung Medical Center. The medical staff attaches the Holter device to the subject and the patch-type electrocardiograph in the Holter examination room. Holter device and patch-type electrocardiograph are returned to the Holter examination room after measuring the electrocardiogram for 24 hours. ECG signals from patch-type electrocardiographs and commercial Holter device and arrhythmia detection information are compared and analyzed.

[Number of Subjects] The number of this study is 105. Written consent is obtained from those who need a Holter examination among outpatients or inpatients at the Pediatrics.

[Data analysis plan] In this study, the personal identification information of the subjects will be coded non-identifying and the collected electrocardiogram data will be stored in the cloud with restricted access from outsiders. Coded, non-identified research data is stored and managed in the cloud in compliance with privacy standards and information security standards. ECG data transmitted through the mobile application is stored in the cloud. Confidentiality and security of data are guaranteed, and data collected from patch-type electrocardiograph is used only for research. The electrocardiogram data acquired in this study are obtained in TXT or MAT format and used for data analysis.

The efficacy evaluation items are as follows.

  1. Comparison of ECG signal concordance

  2. Comparison of the parameters below for a noise-free section of about 1 minute, not the entire section

    • P-wave, QRS-complex, T-wave: Sensitivity, Positive Predictive Value, Amplitude difference, Correlation to the entire signal
    • P, QRS, T (On/Off time): Time difference
    • PR, QRS, QT (Duration/Interval): Time difference
  3. Comparison of arrhythmia diagnosis concordance rates between 24-hour patch-type electrocardiograph and commercial Holter device measurement data by arrhythmia categories (total of 19 including ventricular ectopic beat)

The evaluation method and interpretation method are as follows.

  1. In the case of categorical variables, the sensitivity and positive predictive value are evaluated, and in the case of continuous variables, the concordance between the measured values measured by the patch-type electrocardiograph and the Holter device is evaluated using the Bland-Altman plot.

  2. Amplitude difference of P-wave, QRS-complex and T-wave & Correlation to the entire signal

    • Peaks of P-wave, QRS-complex (R-wave peaks) and peaks of T-wave are manually annotated in the holter device and patch-type ECG independently to assess amplitude difference and correlation to the entire ECG signal.
  3. Time difference of P, QRS, T (On/Off time) & PR, QRS, QT (Duration/Interval)

    • Exact timing of fiducial markers such as onset of P-wave and QRS-complex, and offset of P-wave, QRS-complex, T-wave and clinically useful intervals and durations (PR-interval, RR interval, QT interval, P-wave duration, QRS-duration) are manually annotated in the holter device and patch-type ECG independently to quantify the accuracy of delineating these characteristic points in the patch sensor.
  4. Root Mean Square of Heart Rate

    • Calculate and analyze the square root after averaging the squares of each heart rate value in the analysis section from the electrocardiogram obtained simultaneously from the Holter device and the patch-type electrocardiograph

Enrollment

105 estimated patients

Sex

All

Ages

29 days to 19 years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  • More than 29 days old and less than 19 years old
  • Patients who are need regular Holter electrocardiography among outpatients or hospitalized patients at the Pediatrics, Samsung Medical Center
  • Those who gave written consent to this clinical study or obtained the written consent of the legal representative

Exclusion criteria

  • Patients who cannot give consent through the patient ownself or legal representative
  • If the researcher judges that participation in this study is inappropriate for other reasons

Trial design

Primary purpose

Diagnostic

Allocation

N/A

Interventional model

Single Group Assignment

Masking

None (Open label)

105 participants in 1 patient group

Electrocardiogram measurements by holter device and patch-type electrocardiographic at the same time
Experimental group
Description:
The patient is measured electrocardiogram simultaneously through a holter device and patch-type electrocardiograph
Treatment:
Device: Electrocardiogram measurements by holter device and patch-type electrocardiographic at the same time

Trial contacts and locations