ClinicalTrials.Veeva

Menu

Validation of New G6PD Point of Care Tests Against Gold Standard Quantitative (Biosensors)

University of Oxford logo

University of Oxford

Status

Completed

Conditions

Genotyping
Validation
Biosensors

Study type

Observational

Funder types

Other

Identifiers

NCT02625285
SMRU1501

Details and patient eligibility

About

This is a laboratory research to evaluate performances of quantitative POC G6PD tests (Biosensors) against gold standard quantitative laboratory based tests and genotyping.

Full description

Radical cure of P. vivax malaria can be achieved using large doses of primaquine which have shown to cause some degree of hemolysis even in subjects who tested normal by the conventional qualitative G6PD tests. Different regimens of drug might be more appropriate in subjects with intermediate G6PD activity which can be currently measured only in well equipped laboratories. Validation of new qualitative and quantitative point-of-care tests is essential to for safe deployment of standard and new radical cure regimens against Plasmodium vivax.

This study will evaluate performances of quantitative POC G6PD tests (Biosensors) against gold standard quantitative laboratory based tests and genotyping. Patients will be searched from the electronic record database to identify up to 150 patients with the following characteristics: approximately 50 G6PD-deficient men and women (however only a very small percent of deficients are homozygous women) approximately 50 G6PD-heterozygous women with intermediate activity, and approximately 50 G6PD-normal individuals and patients who meet the screening criteria will be contacted at their home by a clinic staff member. The home visitor will explain study details and invite to participate in the study. If the patients express interest in participating, they will be directed to visit the following month at the clinic where they usually receive care.

After signing the informed consent form, two blood samples will be withdrawn; one sample will be taken by finger-stick (200 µl) and one sample (0.5 ml) will be taken by arm venipuncture. The blood from capillary sample will be used to assess G6PD activity using the POC tests, CBC and laboratory based quantitative tests. The venous blood will be used for reference standard quantitative G6PD assays, Hb typing analysis, CBC and buffy-coat will be stored for DNA extraction for G6PD genotyping only; leftover blood will be discarded.

Enrollment

150 patients

Sex

All

Ages

18+ years old

Volunteers

Accepts Healthy Volunteers

Inclusion criteria

  • Subjects with age ≥ 18 years
  • Subjects willing to participate and sign informed consent form
  • Male and Female
  • Previously tested G6PD deficient, G6PD normal and G6PD intermediate or heterozygous for G6PD variants at SMRU clinic

Exclusion criteria

  • Patients with severe malaria or other severe or any acute illness
  • Patients who received a blood transfusion in the last 3 months
  • Patients who received primaquine in the past 1 month (this is to ensure that previous characterization of phenotype in the healthy subject has not been influenced by a recent hemolytic reaction)
  • Patients who have not had a critical illness or received other hemolytic drugs (this is to ensure that previous characterization of phenotype in the healthy subject has not been influenced by a recent hemolytic reaction)

Trial design

150 participants in 3 patient groups

G6PD Deficient Volunteers
Description:
* Subjects with age ≥ 18 years * Male and female * Previously tested G6PD deficient at SMRU clinic
G6PD Intermediate or Heterozygous Volunteers
Description:
* Subjects with age ≥ 18 years * Female * Previously tested G6PD intermediate or heterozygous for G6PD variants at SMRU clinic
G6PD-Normal Volunteers
Description:
* Subjects with age ≥ 18 years * Male and female * Previously tested G6PD normal at SMRU clinic