The purpose of the study is to:
Full description
Testing will be performed with participants semi-recumbent, under non-motion conditions. SpO2 accuracy of the test device will be determined by a controlled desaturation study with measurements taken over the full range of SaO2 values for which accuracy is to be claimed +3% of the lower value and -3% of the upper value. Participants will be desaturated to achieve six plateaus between 100-70% SaO2.
Arterial blood draws within the hypoxic plateaus will be performed, with draws occurring no less than 30 seconds after reaching a plateau and no less than 20 seconds apart. Five blood draws will be taken at each plateau. Plateaus will be determined using the reference transmittance pulse oximeter and are defined as a change in SpO2 on the reference pulse oximeter of no more than 1% over a duration of 20 seconds. Sampling will not continue if the blood oxygen saturation destabilizes between blood draws.
Time in low O2 saturation decades (i.e., 70-80% and 80-90%) will be limited to the minimum amount of time required to obtain test data and will not exceed 10 minutes. The saturation level at each plateau and the number of plateaus may vary among participants. Arterial pressure, ECG, heart rate, EtCO2, respiratory rate, and FiO2 will be monitored during testing. Extra samples may be collected at the discretion of the PI. The controlled desaturation protocol is expected to take approximately 30 to 60 minutes.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
12 participants in 1 patient group
Loading...
Central trial contact
Bradley Hansen
Data sourced from clinicaltrials.gov
Clinical trials
Research sites
Resources
Legal